DETERMINATION OF TETRACYCLINES IN BOVINE AND PORCINE MUSCLE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING SOLID-PHASE EXTRACTION

被引:74
作者
WALSH, JR
WALKER, LV
WEBBER, JJ
机构
[1] Regional Veterinary Laboratory, Department of Agriculture, Hamilton, Vic. 3300, Ballarat Road
来源
JOURNAL OF CHROMATOGRAPHY | 1992年 / 596卷 / 02期
关键词
D O I
10.1016/0021-9673(92)85009-I
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method is presented for the determination of the three tetracyclines oxytetracycline, tetracycline and chlortetracycline in muscle, spiked at 100 ng/g, using high-performance liquid chromatography (HPLC). The concentration and extraction steps are carried out using Waters Environmental Sep-Pak cartridges. The principal steps involve homogenizing the sample in EDTA-McIlvaine buffer followed by centrifugation and precipitation of the supernatant using trichloroacetic acid. After further filtration and concentration on a Sep-Pak cartridge. the sample is eluted and analysed by HPLC with UV detection and confirmation by diode-array. The column used is a Nova-Pak C18 (4-mu-m) cartridge (10 cm x 8 mm I.D.). A phosphate-citrate acetonitrile buffer, utilizing ion suppression, is the mobile phase. The analytes are detectable at levels down to 10 ng/g. The analyte identity can be confirmed at 20 ng/g by the use of diode-array detection and spectral library comparison.