EFFECT OF INFANT POSITION ON BREATH AMPLITUDE MEASURED BY TRANSTHORACIC IMPEDANCE AND STRAIN-GAUGES

被引:17
作者
BAIRD, TM
NEUMAN, MR
机构
[1] CASE WESTERN RESERVE UNIV,SCH MED,METROHLTH MED CTR,DEPT PEDIAT,CLEVELAND,OH 44106
[2] CASE WESTERN RESERVE UNIV,SCH MED,METROHLTH MED CTR,DEPT OBSTET & GYNECOL,CLEVELAND,OH 44106
关键词
PNEUMOTACHOGRAPHIC TIDAL VOLUME; PRONE; SUPINE; LEFT AND RIGHT LATERAL POSITIONS; CORRELATIONS BETWEEN METHODS; POSITIONS; DIFFERENCES IN MEAN MEDIAN BREATH AMPLITUDES;
D O I
10.1002/ppul.1950100111
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Continuous monitoring of respiration by transthoracic electrical impedance gives a signal that has certain not well understood irregularities. Among them is a change in the amplitude of the signal when there is no apparent change in the infant's tidal respiration. One factor that could hypothetically account for alterations of the impedance signal is a change in current path through the thorax secondary to a change in body position. To test this hypothesis we have studied the relationships between breath amplitude measured by transthoracic impedance, one strain gauge on the chest and one on the abdomen, and tidal volume by integrated flow in four body positions. Median breath amplitude was found to vary significantly with body position according to the measuring device. The median impedance breath amplitude increased by 27% in the supine position compared with the prone position, with no associated change in tidal volume. Differences in the strain gauge signal amplitude for these positions were not statistically significant. Correlation between breath amplitude measured by impedance changes and tidal volume was minimal (r = 0.114). These results indicate that infant position affects impedance breath amplitude independently of changes in tidal volume.
引用
收藏
页码:52 / 56
页数:5
相关论文
共 12 条
[1]   VOLUMETRIC DYNAMICS OF RESPIRATION AS MEASURED BY ELECTRICAL IMPEDANCE PLETHYSMOGRAPHY [J].
ALLISON, RD ;
HOLMES, EL ;
NYBOER, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1964, 19 (01) :166-&
[2]  
BANCALARI E, 1986, NEONATAL PULMONARY C, P205
[3]   CHANGES IN RESPIRATORY PATTERN RESULTING FROM THE USE OF A FACEMASK TO RECORD RESPIRATION IN NEWBORN-INFANTS [J].
FLEMING, PJ ;
LEVINE, MR ;
GONCALVES, A .
PEDIATRIC RESEARCH, 1982, 16 (12) :1031-1034
[4]   IMPEDANCE MEASUREMENT OF TIDAL VOLUME AND VENTILATION [J].
HAMILTON, LH ;
BEARD, JD ;
KORY, RC .
JOURNAL OF APPLIED PHYSIOLOGY, 1965, 20 (03) :565-&
[5]  
NEUMAN MR, 1984, MED TECHNOLOGY NEONA, P49
[6]  
NEUMAN MR, 1987, NIH872905 PUBL
[7]  
OLSSON T, 1970, ACTA PEDIATR SCAND S, V207, P1
[8]  
POLGAR G, 1965, PEDIATRICS, V36, P861
[9]   IMPROVED OXYGENATION AND LUNG COMPLIANCE WITH PRONE POSITIONING OF NEONATES [J].
WAGAMAN, MJ ;
SHUTACK, JG ;
MOOMJIAN, AS ;
SCHWARTZ, JG ;
SHAFFER, TH ;
FOX, WW .
JOURNAL OF PEDIATRICS, 1979, 94 (05) :787-791
[10]  
WHITNEY RJ, 1953, J PHYSIOL-LONDON, V121, P1