IMPROVING THE SIGNAL-TO-NOISE RATIO IN MASS AND ION KINETIC-ENERGY SPECTROMETERS

被引:5
作者
BRENTON, AG
BEYNON, JH
MORGAN, RP
机构
[1] Royal Society Research Unit Uniuersity College of Swansea Singleton Park, Swansea
来源
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES | 1979年 / 31卷 / 1-2期
关键词
D O I
10.1016/0020-7381(79)80103-5
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The signal-to-noise ratio in mass and ion kinetic energy spectrometers is limited by noise generated from the presence of scattered ions and neutrals. Methods of eliminating this are illustrated with reference to the ZAB-2F instrument manufactured by VG-Micromass Ltd. It is estimated that after the modifications the instrument is capable on a routine basis of measuring peaks corresponding to the arrival of ions at a rate of the order of 1 ion s-l. © 1979.