MONITORING OF THE X-RAY-BEAM INTENSITY GOING THROUGH A COLLIMATOR USING A SMALL IONIZATION-CHAMBER

被引:3
作者
BOSSHARD, R
KAHN, R
FOURME, R
机构
[1] LURE, Bat. 209 D, Université Paris-Sud, 91405, Orsay
关键词
D O I
10.1063/1.1142616
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe a device which monitors the intensity of a collimated monochromatic x-ray beam. A small ionization chamber, mounted between the collimator and the sample, is connected to a current integrator. It is constructed and operated in order to have a linear response. Technical details about electronics are given. Two diffractometers using synchrotron radiation from bending magnets are equipped with this device, with the main purpose of rescaling diffraction data from macromolecular crystals.
引用
收藏
页码:810 / 813
页数:4
相关论文
共 12 条
[1]   THE SATURATION CURVE AT HIGH IONIZATION INTENSITY [J].
BOAG, JW ;
WILSON, T .
BRITISH JOURNAL OF APPLIED PHYSICS, 1952, 3 (JUL) :222-229
[2]  
BOSSHARD R, 1992, REV SCI INSTRUM, V63
[3]   STABILIZATION OF SYNCHROTRON RADIATION BEAM AT HASYLAB [J].
BREFELD, W .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1513-1516
[4]   PICOAMMETER FOR ION-CHAMBER SPANS 6 DECADES WITHOUT RANGE SWITCHING [J].
BURKHARDT, W ;
IACOPINI, M ;
MARANESI, P .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (01) :311-313
[5]   CHARGE-BALANCING CURRENT INTEGRATOR WITH LARGE DYNAMIC-RANGE [J].
GOTTSCHALK, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 207 (03) :417-421
[6]   REVIEW OF SYNCHROTRON BEAM STABILITY AND STABILIZING SYSTEMS [J].
HETTEL, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1501-1506
[7]  
HINE G, 1956, RAD DOSIMETRY, P164
[8]  
KAHN R, 1982, NUCL INSTRUM METHODS, V201, P203, DOI 10.1016/0167-5087(82)90028-X
[9]  
KAHN R, 1984, UNPUB 13TH C INT UN
[10]   PRECISE AND RAPID MEASUREMENTS OF SMALL CURRENTS FROM HIGH IMPEDANCE SOURCES [J].
SAMUELSON, G ;
BENGTSSON, LG .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1973, 44 (07) :920-921