SEMIANALYTICAL EXPRESSIONS FOR (LBAR/P)(AIR)(MED) AND PREPL FOR ELECTRON-BEAMS

被引:1
作者
WIELOPOLSKI, L
PAI, S
MLYN, M
机构
关键词
TG-21; STOPPING POWER; FLUENCE CORRECTION; IONIZATION DOSE; CONVERSION;
D O I
10.1118/1.596745
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The tables of the mean restricted collision mass stopping power ratios for water, polystyrene and acrylic relative to air given in the AAPM TG-21 protocol have been fitted to an expression with 20 coefficients using the depth in the phantom and the mean incident electron energy as two independent variables. Using these expressions, the calculated values agree with the tabulated values within +0.5% in 85% of the cases and within +/-1.0% in 95% of the cases. For each of the four cylindrical chamber inner diameters, given in the protocol, the electron fluence correction P(repl) has been fitted to an expression with four coefficients using the mean electron energy at depth z as an independent variable.