PATIENT DOSE AND RISK IN MAMMOGRAPHY

被引:15
作者
LAW, J [1 ]
机构
[1] WESTERN GEN HOSP,EDINBURGH EH4 2XU,MIDLOTHIAN,SCOTLAND
关键词
DOSE; RISK; MAMMOGRAPHY; BREAST SCREENING;
D O I
10.1259/0007-1285-64-760-360
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Patient dose in mammography is estimated by two different methods which are compared and which give good agreement. A mean tissue dose of about 1 mGy per film is found for a breast of 4.5 cm compressed thickness. Variables which affect dose are then considered quantitatively, including compressed breast thickness, tube potential, grids, magnification and beam collimation. The variables having the greatest effect are breast thickness and magnification. For example, an 8 cm breast receives about four times the mean tissue dose of a 3.5 cm breast. Similarly, using a magnification factor of about 2, dose is increased about four times in the primary beam, but the adverse consequences may be largely removed with conventional collimation to part of the breast only. Finally, the dose estimates are combined with existing data on breast cancer induction to predict the risk of carcinogenesis in a breast screening programme. For example, in a screening centre performing 15 000 examinations per year, only one induced cancer is predicted in about 7 years of screening under average UK conditions of age and breast thickness.
引用
收藏
页码:360 / 365
页数:6
相关论文
共 8 条
[1]   ENHANCED RISK FROM LOW-ENERGY SCREEN FILM MAMMOGRAPHY X-RAYS [J].
BRENNER, DJ ;
AMOLS, HI .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (742) :910-914
[2]   FURTHER COMPARISONS OF FILMS, SCREENS AND CASSETTES FOR MAMMOGRAPHY [J].
LAW, J ;
KIRKPATRICK, AE .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (746) :128-131
[3]   CANCERS INDUCED AND CANCERS DETECTED IN A MAMMOGRAPHY SCREENING-PROGRAM [J].
LAW, J .
BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (711) :231-234
[4]  
LAW J, 1988, ARE XRAYS SAFE ENOUG, P78
[5]   MAMMOGRAPHY AND SCREENING FOR BREAST-CANCER [J].
ROEBUCK, EJ .
BRITISH MEDICAL JOURNAL, 1986, 292 (6515) :223-226
[6]  
SHIMUZU Y, 1987, TR1287 RERF RAD EFF
[7]   DOSES TO PATIENTS FROM ROUTINE DIAGNOSTIC-X-RAY EXAMINATIONS IN ENGLAND [J].
SHRIMPTON, PC ;
WALL, BF ;
JONES, DG ;
FISHER, ES ;
HILLIER, MC ;
KENDALL, GM ;
HARRISON, IRM .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (704) :749-758
[8]  
1989, COMMISSIONING ROUTIN