ELECTROCARDIOGRAM ACQUISITION DURING MR EXAMINATIONS FOR PATIENT MONITORING AND SEQUENCE TRIGGERING

被引:47
作者
FELBLINGER, J [1 ]
LEHMANN, C [1 ]
BOESCH, C [1 ]
机构
[1] UNIV BERN,DEPT ANAT,CH-3000 BERN,SWITZERLAND
关键词
ECG; MAGNETIC RESONANCE; PATIENT MONITORING; GATING/TRIGGERING;
D O I
10.1002/mrm.1910320416
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have developed a method for measuring the electrocardiogram (ECG) continuously during MR examinations. In contrast to ECG acquisition by wires, our new method is to amplify and convert the ECG into an optical signal directly above the patient's heart. The optical signal is transmitted out of the magnet bore by optical fiber. The small and fixed dimensions of the ECG amplifier avoid interactions with the MR system because of shorter electrical structures and smaller enclosed areas. Tests of the proposed device in a 1.5 Tesla MR system show that continuous and reliable ECG monitoring and sequence triggering are possible.
引用
收藏
页码:523 / 529
页数:7
相关论文
共 19 条
[1]   BURNS ASSOCIATED WITH PULSE OXIMETRY DURING MAGNETIC-RESONANCE-IMAGING [J].
BASHEIN, G ;
SYROVY, G .
ANESTHESIOLOGY, 1991, 75 (02) :382-383
[2]  
BLAKELEY DM, 1985, 4TH P SMRM ANN M LON, P1073
[3]  
BOESCH C, 1989, 8TH P SMRM ANN M AMS, P999
[4]  
Buchwald Randall H., 1990, United States Patent, Patent No. [US 4951672 A, 4951672]
[5]   RF INTERFERENCE SUPPRESSION IN A CARDIAC SYNCHRONIZATION SYSTEM OPERATING IN A HIGH MAGNETIC-FIELD NMR IMAGING-SYSTEM [J].
DAMJI, AA ;
SNYDER, RE ;
ELLINGER, DC ;
WITKOWSKI, FX ;
ALLEN, PS .
MAGNETIC RESONANCE IMAGING, 1988, 6 (06) :637-640
[6]   OPTIMIZING ELECTROCARDIOGRAPH ELECTRODE PLACEMENT FOR CARDIAC-GATED MAGNETIC-RESONANCE-IMAGING [J].
DIMICK, RN ;
HEDLUND, LW ;
HERFKENS, RJ ;
FRAM, EK ;
UTZ, J .
INVESTIGATIVE RADIOLOGY, 1987, 22 (01) :17-22
[7]  
FELBLINGER J, 1993, 13TH P SMRM ANN M NE, P1374
[8]  
FELBLINGER J, 1990, THESIS INPL FRANCE
[9]  
FELBLINGER J, 1993, 15TH P IEEE EMBS ANN, P191
[10]  
FELBLINGER J, 1993, Patent No. 9304918