ABSOLUTE CALIBRATION OF TOKAMAK FUSION TEST REACTOR NEUTRON DETECTORS FOR D-T PLASMA OPERATION

被引:23
作者
JASSBY, DL [1 ]
BARNES, CW [1 ]
JOHNSON, LC [1 ]
ROQUEMORE, AL [1 ]
STRACHAN, JD [1 ]
JOHNSON, DW [1 ]
MEDLEY, SS [1 ]
YOUNG, KM [1 ]
机构
[1] LOS ALAMOS NATL LAB,LOS ALAMOS,NM
关键词
D O I
10.1063/1.1146196
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The two most sensitive TFTR fission-chamber detectors were absolutely calibrated in situ by a D-T neutron generator (∼5×107 n/s) rotated once around the torus in each direction, with data taken at about 45 positions. The combined uncertainty for determining fusion neutron rates, including the uncertainty in the total neutron generator output (±9%), counting statistics, the effect of coil coolant, detector stability, cross calibration to the current mode or log Campbell mode and to other fission chambers, and plasma position variation, is about ±13%. The NE-451 (ZnS) scintillators and 4He proportional counters that view the plasma in up to 10 collimated sightlines were calibrated by scanning the neutron generator radially and toroidally in the horizontal midplane across the flight tubes of 7 cm diam. Spatial integration of the detector responses using the calibrated signal per unit chord-integrated neutron emission gives the global neutron source strength with an overall uncertainty of ±14% for the scintillators and ±15% for the 4He counters. © 1995 American Institute of Physics.
引用
收藏
页码:891 / 893
页数:3
相关论文
共 13 条
  • [1] NEUTRON DIAGNOSTICS ON TFTR UTILIZING THE CAMPBELLING TECHNIQUE
    ENGLAND, AC
    HENDEL, HW
    NIESCHMIDT, EB
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1986, 57 (08) : 1754 - 1756
  • [2] TFTR EPITHERMAL NEUTRON DETECTOR SYSTEM - RECALIBRATION AND EFFECT OF NONISOTROPIC NEUTRON EMISSION
    HENDEL, HW
    JASSBY, DL
    BOSCH, HS
    BARNES, CW
    JOHNSON, LC
    MURPHY, TJ
    NIESCHMIDT, EB
    SAITO, T
    STRACHAN, JD
    TAIT, GD
    YOUNG, KM
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (08) : 1682 - 1684
  • [3] INSITU CALIBRATION OF TFTR NEUTRON DETECTORS
    HENDEL, HW
    PALLADINO, RW
    BARNES, CW
    DIESSO, M
    FELT, JS
    JASSBY, DL
    JOHNSON, LC
    KU, LP
    LIU, QP
    MOTLEY, RW
    MURPHY, HB
    MURPHY, J
    NIESCHMIDT, EB
    ROBERTS, JA
    SAITO, T
    STRACHAN, JD
    WASZAZAK, RJ
    YOUNG, KM
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (07) : 1900 - 1914
  • [4] RELATIVE INTENSITIES OF 2.5-MEV AND 14-MEV SOURCE NEUTRONS FROM COMPARATIVE RESPONSES OF U-238 AND U-235 DETECTORS
    JASSBY, DL
    HENDEL, HW
    BOSCH, HS
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (08) : 1688 - 1690
  • [6] CROSS CALIBRATION OF NEUTRON DETECTORS FOR DEUTERIUM-TRITIUM OPERATION IN TFTR
    JOHNSON, LC
    BARNES, CW
    DUONG, HH
    HEIDBRINK, WW
    JASSBY, DL
    LOUGHLIN, MJ
    ROQUEMORE, AL
    RUSKOV, E
    STRACHAN, JD
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01) : 894 - 896
  • [7] CALCULATION OF THE ABSOLUTE DETECTION EFFICIENCY OF A MODERATED U-235 NEUTRON DETECTOR ON THE TOKAMAK FUSION TEST REACTOR
    KU, LP
    HENDEL, HW
    LIEW, SL
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 280 (01) : 113 - 122
  • [8] MEASUREMENT OF DT NEUTRON EMISSION FROM TFTR WITH HE-4 PROPORTIONAL RECOIL COUNTERS
    MCCAULEY, JS
    STRACHAN, JD
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) : 4536 - 4538
  • [9] EFFECTS OF NEUTRON ENERGY-SPECTRUM ON THE EFFICIENCY CALIBRATION OF EPITHERMAL NEUTRON DETECTORS
    NIESCHMIDT, EB
    ENGLAND, AC
    HENDEL, HW
    HILLIS, DL
    ISAACSON, JA
    KU, LP
    TSANG, FY
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1985, 56 (05) : 1084 - 1086
  • [10] TFTR MULTICHANNEL NEUTRON COLLIMATOR
    ROQUEMORE, AL
    CHOUINARD, RC
    DIESSO, M
    PALLADINO, R
    STRACHAN, JD
    TAIT, GD
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) : 3163 - 3165