DETERMINATION OF MOLYBDENUM IN VARIOUS MATERIALS BY NORMAL-PHASE LIQUID-CHROMATOGRAPHY USING N-BENZOYL-N-PHENYLHYDROXYLAMINE

被引:11
作者
BAGUR, G [1 ]
SANCHEZVINAS, M [1 ]
GAZQUEZ, D [1 ]
机构
[1] UNIV GRANADA,FAC SCI,DEPT ANALYT CHEM,E-18071 GRANADA,SPAIN
关键词
LIQUID CHROMATOGRAPHY; MOLYBDENUM; N-BENZOYL-N-PHENYLHYDROXYLAMINE; EXTRACTION; FOODS; FERTILIZERS; VITAMIN COMPLEX;
D O I
10.1016/0003-2670(95)00081-A
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学]; 081704 [应用化学];
摘要
A normal-phase liquid chromatographic method for the selective determination of molybdenum with N-benzoyl-N-phenylhydroxylamine is described. The molybdenum(VI) complex was preconcentrated by extraction into chloroform and injected onto a nitrile column for chromatography. The mobile phase was a 0.075 M solution of reagent in chloroform (stabilized with amylene). The detection limit for molybdenum by the proposed method was 0.88 ng for a phase volume ratio of 20:1 (aqueous to organic). Molybdenum has been determined in several samples with satisfactory accuracy and precision.
引用
收藏
页码:157 / 162
页数:6
相关论文
共 9 条
[1]
REVERSED-PHASE HPLC SEPARATION OF COMPLEX-MIXTURES OF TRACE-METALS AS DIBENZYLDITHIOCARBAMATE CHELATES [J].
BAIOCCHI, C ;
MARCHETTO, A ;
SAINI, G ;
BERTOLO, P .
TALANTA, 1988, 35 (09) :685-691
[2]
EXTRACTION AND SPECTROPHOTOMETRIC DETERMINATION OF MOLYBDENUM(VI) WITH N-PHENYLBENZOHYDROXAMIC ACID AND PHENYLFLUORONE [J].
GUNAWARDHANA, HD ;
FERNANDO, SA .
CANADIAN JOURNAL OF CHEMISTRY-REVUE CANADIENNE DE CHIMIE, 1987, 65 (05) :1124-1127
[3]
SIMULTANEOUS DETERMINATION OF COPPER AND IRON IN BOVINE LIVER AND TOMATO LEAVES BY REVERSED PHASE HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY [J].
ICHINOKI, S ;
YAMAZAKI, M .
ANALYTICAL SCIENCES, 1989, 5 (04) :465-469
[4]
MAJUNDAR AK, 1972, N BPHA ITS ANALOGUES
[5]
HIGH-PRESSURE LIQUID-CHROMATOGRAPHY OF METAL-CHELATES OF DIETHYLDITHIOCARBAMIC ACID AND DITHIZONE [J].
OLAUGHLIN, JW ;
OBRIEN, TP .
ANALYTICAL LETTERS, 1978, 11 (10) :829-844
[6]
Reilly C., 1980, METAL CONTAMINATION
[7]
RODRIGUEZ LC, 1995, J ASSOC OFF ANA CHEM, V78, P795
[8]
Snyder L.R., 2012, PRACTICAL HPLC METHO
[9]
AN ION-PAIR REVERSED-PHASE HPLC FLUORIMETRIC SYSTEM FOR ULTRATRACE DETERMINATION OF ALUMINUM WITH SALICYLALDEHYDEBENZOYLHYDRAZONE [J].
UEHARA, N ;
KANBAYASHI, M ;
HOSHINO, H ;
YOTSUYANAGI, T .
TALANTA, 1989, 36 (10) :1031-1035