PROTEIN LAYER ON HEMODIALYSIS MEMBRANES - A NEW IMMUNOHISTOCHEMISTRY TECHNIQUE

被引:12
作者
PANICHI, V
CASAROSA, L
GATTAI, V
BIANCHI, AM
ANDREINI, B
MIGLIORI, M
DEPIETRO, S
GIOVANNINI, L
PALLA, R
机构
[1] UNIV PISA,INST MED CLIN 2,I-56100 PISA,ITALY
[2] CNR,I-56100 PISA,ITALY
[3] UNIV PISA,INST PHARMACOL,PISA,ITALY
关键词
PROTEIN LAYER; BIOCOMPATIBILITY; IMMUNOHISTOCHEMISTRY;
D O I
10.1177/039139889501800602
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to investigate the nature of the protein layer deposited on hemodialysis membranes we developed a direct immunohistochemical technique using fluoresceinated antibodies to plasma proteins. Fifteen patients on regular dialytic treatment were dialyzed with CU, HE, PAN, PS and PMIMA and the dialyzers analyzed at the end of a standard dialytic session. Snap frozen sections of hollow fiber devices were treated with flourescein-isothiocyanate conjugated antibodies for IgG, IgA, IgM, C3c, fibrinogen, factor VIII, factor XIIIa-s, antithrombin III, fibrinogen degradation products (PDF) and plasminogen. Protein deposits were evaulated by a quantitative criteria, which evaluates the intensity of fluorescence and the area of the capillary wall occupied by this fluorescence by using an image analysis software. The coagulation cascade is activated by all membranes and similar deposits of these proteins were revealed, whereas important differences in C3c deposition were found.
引用
收藏
页码:305 / 308
页数:4
相关论文
共 9 条
[1]  
ANDERSON JM, 1990, INT J ARTIF ORGANS, V13, P357
[2]  
Baier R E, 1969, J Biomed Mater Res, V3, P191, DOI 10.1002/jbm.820030115
[3]  
BRASH JL, 1988, BLOOD, V71, P932
[4]   ADHERENCE OF NEUTROPHILS TO HEMODIALYSIS MEMBRANES - ROLE OF COMPLEMENT RECEPTORS [J].
CHEUNG, AK ;
HOHNHOLT, M ;
GILSON, J .
KIDNEY INTERNATIONAL, 1991, 40 (06) :1123-1133
[5]  
COLTON CK, 1994, NEPHROL DIAL TRANSPL, V9, P11
[6]  
KUWAHARA T, 1989, ARTIF ORGANS, V5, P427
[7]  
LANE DA, 1994, NEPHROL DIAL TRANSPL, V9, P18
[8]  
PANICHI V, 1988, IMPLANTS MATERIALS B, P193
[9]  
VROMAN L, 1988, B NEW YORK ACAD MED, V64, P352