GEOMETRY DEPENDENCE OF THE EMPIRICAL RELATIONSHIP BETWEEN VOLUME AND EFFICIENCY FOR GE DETECTORS

被引:1
作者
PASCARELLE, SM [1 ]
OWENS, A [1 ]
GEHRELS, N [1 ]
TEEGARDEN, BJ [1 ]
机构
[1] NASA,GODDARD SPACE FLIGHT CTR,HIGH ENERGY ASTROPHYS LAB,GREENBELT,MD 20771
关键词
D O I
10.1016/0168-9002(91)90429-T
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have tested the suggestion of Zhong that the slope of the efficiency curve is nearly independent of the angle of incidence of the illuminating source and the detector height-to-diameter ratio. Within the chosen range of energies and ratios, the slope is found to be constant within approximately 3%, but for larger ranges, it may vary by as much as 10% as a function of angle of incidence and 56% as a function of height-to-diameter ratio.