ACOUSTIC POWER CALIBRATIONS OF CYLINDRICAL INTRACAVITARY ULTRASOUND HYPERTHERMIA APPLICATORS

被引:69
作者
HYNYNEN, K [1 ]
机构
[1] UNIV ARIZONA,HLTH SCI CTR,DEPT RADIAT ONCOL,TUCSON,AZ 85724
关键词
HYPERTHERMIA; ULTRASOUND; INTRACAVITARY; QUALITY ASSURANCE;
D O I
10.1118/1.597094
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Preliminary clinical results indicate that some tumors can be heated well utilizing cylindrical ultrasound sources placed in body cavities. In this paper a simple method for measuring the acoustic power from cylindrical intracavitary transducers will be described. The radially propagating acoustic field was converted to a beam with a single propagation direction by a brass reflector, and the radiation force generated by this beam on an absorbing target was measured. The power output of several clinical intracavitary arrays varied significantly between identically shaped transducer elements. The results show that it is important to measure the acoustic power output from each element prior to its clinical use. The radiation force technique is simple and sensitive and can be easily adapted to be used as a routine clinical quality assurance method.
引用
收藏
页码:129 / 134
页数:6
相关论文
共 16 条
[1]  
ANHALT DP, 1988, P ANN INT C IEEE ENG, V10, P876
[2]   RADIATION PRESSURE IN A SOUND WAVE [J].
BEYER, RT .
AMERICAN JOURNAL OF PHYSICS, 1950, 18 (01) :25-29
[3]   ACOUSTIC RADIATION PRESSURE OF PLANE COMPRESSIONAL WAVES [J].
BORGNIS, FE .
REVIEWS OF MODERN PHYSICS, 1953, 25 (03) :653-664
[4]   SPEED OF SOUND IN PURE WATER [J].
DELGROSS.VA ;
MADER, CW .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1972, 52 (05) :1442-&
[5]  
DIEDERICH C, 1987, P IEEE ULTRASON S, V20, P871
[6]   THE DEVELOPMENT OF INTRACAVITARY ULTRASONIC APPLICATORS FOR HYPERTHERMIA - A DESIGN AND EXPERIMENTAL-STUDY [J].
DIEDERICH, CJ ;
HYNYNEN, K .
MEDICAL PHYSICS, 1990, 17 (04) :626-634
[7]   INDUCTION OF HYPERTHERMIA USING AN INTRACAVITARY MULTIELEMENT ULTRASONIC APPLICATOR [J].
DIEDERICH, CJ ;
HYNYNEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (04) :432-438
[8]   THE FEASIBILITY OF USING ELECTRICALLY FOCUSED ULTRASOUND ARRAYS TO INDUCE DEEP HYPERTHERMIA VIA BODY CAVITIES [J].
DIEDERICH, CJ ;
HYNYNEN, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1991, 38 (03) :207-219
[9]  
FOSMIRE H, IN PRESS INT J RAD O
[10]  
GAUTHERIE M, 1990, CLIN THERMOTHERAPY