PENALIZED WEIGHTED LEAST-SQUARES IMAGE-RECONSTRUCTION FOR POSITRON EMISSION TOMOGRAPHY

被引:443
作者
FESSLER, JA
机构
[1] University of Michigan Medical Center, Ann Arbor
基金
美国国家卫生研究院;
关键词
D O I
10.1109/42.293921
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper presents an image reconstruction method for positron-emission tomography (PET) based on a penalized, weighted least-squares (PWLS) objective. For PET measurements that are precorrected for accidental coincidences, we argue statistically that a least-squares objective function is as appropriate, if not more so, than the popular Poisson likelihood objective. We propose a simple data-based method for determining the weights that accounts for attenuation and detector efficiency. A non-negative successive over-relaxation (+SOR) algorithm converges rapidly to the global minimum of the PWLS objective. Quantitative simulation results demonstrate that the bias/variance trade-off of the PWLS+SOR method is comparable to the maximum-likelihood expectation-maximization (ML-EM) method (but with fewer iterations), and is improved relative to the conventional filtered backprojection (FBP) method. Qualitative results suggest that the streak artifacts common to the FBP method are nearly eliminated by the PWLS+SOR method, and indicate that the proposed method for weighting the measurements is a significant factor in the improvement over FBP.