APPLICATION OF THE SYVA EMIT AND ABBOTT TDX AMPHETAMINE IMMUNOASSAYS TO THE DETECTION OF 3,4-METHYLENE-DIOXYMETHAMPHETAMINE (MDMA) AND 3,4-METHYLENE-DIOXYETHAMPHETAMINE (MDEA) IN URINE

被引:28
作者
KUNSMAN, GW [1 ]
MANNO, JE [1 ]
COCKERHAM, KR [1 ]
MANNO, BR [1 ]
机构
[1] LOUISIANA STATE UNIV,DEPT PHARMACOL,TOXICOL SECT,SHREVEPORT,LA 71130
关键词
D O I
10.1093/jat/14.3.149
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
MDMA and MDEA are hallucinogenic analogs of amphetamine. The need for laboratory monitoring of these substances has developed as a result of their increased recreational use. Since the Abbott TDx®and Syva EMIT®-d.a.u.™ immunoassays are commonly used tests for urine monitoring of drugs-of-abuse, the amphetamine assay of each manufacturer was assessed to determine the degree of cross-reactivity with MDMA and MDEA. Cross-reactivity was evaluated using a series of MDMA- and MDEA-spiked urine samples. Testing was performed over a two-day period with 3 runs/day and each sample run in duplicate. The Syva EMIT d.a.u. amphetamine assay was positive only at the highest concentration standard (10.0 μg/mL) for both MDMA and MDEA. Consequently, no further testing was performed with this assay. Calibration curves were generated for the TDx runs and percent cross-reactivity determinations were made. Precision for the TDx data was evaluated based on within-day and between-day coefficients of variation (CV). CVs for MDMA runs were below 6% for within-day and below 5% for between-day runs. Values of CV for MDEA were below 16% for both within-day and between-day runs; CVs were less than 2.5% for positive values. Cross-reactivity for MDMA ranged from 18% (10.00 μg/mL) to 118% (0.15 μg/mL). Cross-reactivity for the MDEA standards ranged from 12% (10.00 μg/mL) to 47% (0.15 μg/mL). The presence of MDMA and/or MDEA in samples resulting in a negative EMIT-d.a.u. test and a positive TDx test was confirmed by GC/MS analysis. The urine sample was extracted and then derivatized with heptafluorobutyric anhydride prior to analysis with a Hewlett-Packard Model 5988A gas chromatograph/mass spectrometer using the electron impact-selected ion monitoring mode. © 1990 Oxford University Press.
引用
收藏
页码:149 / 153
页数:5
相关论文
共 19 条
  • [1] BAUM RM, 1985, CHEM ENG NEWS 0909, P7
  • [2] CENTRALLY ACTIVE N-SUBSTITUTED ANALOGS OF 3,4-METHYLENEDIOXYPHENYLISOPROPYLAMINE (3,4-METHYLENEDIOXYAMPHETAMINE)
    BRAUN, U
    SHULGIN, AT
    BRAUN, G
    [J]. JOURNAL OF PHARMACEUTICAL SCIENCES, 1980, 69 (02) : 192 - 195
  • [3] MULTIPLE SEVERE COMPLICATIONS FROM RECREATIONAL INGESTION OF MDMA (ECSTASY)
    BROWN, C
    OSTERLOH, J
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1987, 258 (06): : 780 - 781
  • [4] CLARK CC, 1984, J FORENSIC SCI, V29, P1056
  • [5] EVE AND ECSTASY - A REPORT OF 5 DEATHS ASSOCIATED WITH THE USE OF MDEA AND MDMA
    DOWLING, GP
    MCDONOUGH, ET
    BOST, RO
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1987, 257 (12): : 1615 - 1617
  • [6] EISNER B, 1988, HIGH TIMES SEP, P75
  • [7] EISNER B, 1988, HIGH TIMES OCT, P64
  • [8] EISNER B, 1988, HIGH TIMES SEP, P77
  • [9] EISNER B, 1988, HIGH TIMES SEP, P71
  • [10] EISNER B, 1988, HIGH TIMES AUG, P32