NEW TRANSDUCER FOR RESPIRATORY MONITORING - DESCRIPTION OF A HOT-WIRE ANEMOMETER AND A TEST PROCEDURE FOR GENERAL USE

被引:4
作者
KANN, T [1 ]
HALD, A [1 ]
JORGENSEN, FE [1 ]
机构
[1] DISA ELEKTR,SKOVLUNDE,DENMARK
关键词
D O I
10.1111/j.1399-6576.1979.tb01462.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A respiratory transducer based on the constant temperature anemometry principle has been developed for respiratory monitoring, and as a tool for bedside evaluation of pulmonary function. The transducer is characterized by a dynamic range from 0–2.5 l/s and an upper limiting frequency of 50 Hz. It is designed with a view to a low pressure drop of 2.5 mbar/l/s and a minimum dead space of 5 ml. The transducer has been tested using a generally applicable procedure which includes both static and dynamic test set‐ups. The influence of variations in gas composition, temperature and pressure, together with variations in tidal volume and respiration rate, have been investigated. The results show that the transducer registers the immediate value of gas flow‐rate with a mean error less than 5–10% in all situations which are predictable in clinical use. The mean error can be reduced to less than 5% when systematic errors are eliminated. © 1979 The Scandinavian Society of Anaesthesiologists
引用
收藏
页码:349 / 358
页数:10
相关论文
共 6 条
[1]  
APPEL E, 1976, ELEKTRONISCHE BESTIM
[2]  
FRANETZSKI M, 1975, THESIS TH KARLSRUHE
[3]  
HILL DW, 1973, ELECTRONIC TECHNIQUE
[4]  
KING LV, 1914, PHIL T ROY SOC LOND, V24, P520
[5]  
RASMUSSEN CC, 1969, HOT WIRE HOT FILM AN
[6]  
RASMUSSEN CG, 1969, FORSKNING 4, P102