THE SINGLE-BREATH TEST IN NEONATES - DOES PRESSURIZATION OF THE PNEUMOTACHOGRAPH MAKE A DIFFERENCE

被引:2
作者
COATES, AL [1 ]
FLETCHER, ME [1 ]
DUNDAS, I [1 ]
STOCKS, J [1 ]
机构
[1] INST CHILD HLTH,PORTEX DEPT PAEDIAT ANAESTHESIA,LONDON,ENGLAND
关键词
PNEUMOTACHOGRAPHS; INFANT LUNG FUNCTION TESTING; SINGLE BREATH TEST;
D O I
10.1002/ppul.1950190509
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The single breath test for the measurement of respiratory system resistance and compliance in newborns consists of an end inspiratory occlusion which is subsequently released, allowing expiration to proceed through a pneumotachograph (PNT). The measured flow is then integrated to give volume. The simplicity of the test is one of the major reasons for its popularity. However, some investigators have cautioned against the use of an occlusion distal to the PNT because pressurization of the PNT may introduce artifacts in the flow measurement. Despite this caution, many commercial systems use a pressurized PNT. This study investigated the errors that would result from pressurization of the PNT by providing a step function of flow to two infant PNTs, a Fleisch #0 and a Hans Rudolph 4500, in the unpressurized and pressurized state. In each case there was an initial rapid rise of the flow signal, followed by some overshoot and oscillations that rapidly died out. The overshoot and oscillations for the Hans Rudolph PNT were greater when pressurized whereas pressurization had little effect on the Fleisch PNT. Unpressurized, the two were similar. In either case, the artifact introduced by pressurization of the PNT died out so quickly that it would have little effect on any measurement in an infant. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:299 / 306
页数:8
相关论文
共 8 条
[1]  
BATES JHT, 1988, PROD CAN MED BIOL EN, P25
[2]   LINEARITY AND FREQUENCY-RESPONSE OF PNEUMOTACHOGRAPHS [J].
FINUCANE, KE ;
DAWSON, SV ;
EGAN, BA .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 32 (01) :121-&
[3]   RESPIRATORY COMPLIANCE IN INFANTS - A PRELIMINARY EVALUATION OF THE MULTIPLE INTERRUPTER TECHNIQUE [J].
FLETCHER, ME ;
DEZATEUX, CA ;
STOCKS, J .
PEDIATRIC PULMONOLOGY, 1992, 14 (02) :118-125
[4]  
JACKSON AC, 1984, TECHNIQUES LIFE SCI, P1
[5]  
LESOUEF PN, 1984, AM REV RESPIR DIS, V129, P552
[6]   HARMONIC CONTENT OF CERTAIN RESPIRATORY FLOW PHENOMENA OF NORMAL INDIVIDUALS [J].
MCCALL, CB ;
HYATT, RE ;
NOBLE, FW ;
FRY, DL .
JOURNAL OF APPLIED PHYSIOLOGY, 1957, 10 (02) :215-218
[7]  
VANVALKENBURG ME, 1964, NETWORK ANALYSIS, P342
[8]   SINGLE-BREATH METHOD FOR MEASUREMENT OF RESPIRATORY MECHANICS IN ANESTHETIZED ANIMALS [J].
ZIN, WA ;
PENGELLY, LD ;
MILICEMILI, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1982, 52 (05) :1266-1271