EFFECT OF HEAVY-METAL IONS ON THE EFFICACY OF A MIXED BACILLI BOD SENSOR

被引:26
作者
LI, F [1 ]
TAN, TC [1 ]
机构
[1] NATL UNIV SINGAPORE,DEPT CHEM ENGN,SINGAPORE 0511,SINGAPORE
关键词
BOD SENSOR; HEAVY METAL IONS EFFECT; BACILLI CULTURE;
D O I
10.1016/0956-5663(94)80030-8
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Heavy metal ions such as cadmium, mercuric, lead, chromic, stannous, ferrous, ferric and aluminium at a concentration of 4 mM have negligible or no effect on the BOD sensing of a mixed Bacillus subtilis and Bacillus licheniformis 7B microbial BOD sensor. In the presence of cobalt, zinc and nickel ions, the BOD measured by the sensor was higher than the true value of the test solution. This was attributed to the increase in the microbial activity by the metal ions present. Cupric and in particular silver ions significantly suppressed the microbial assimilation of the substrate resulting in a lower measured BOD value of the substrate. The interference of silver and cupric ions on the BOD measurement were successfully eliminated by chelating these ions with EDTA or by phenotypic modification of the microbial system by pre-conditioning the sensor in a BOD solution containing the interfering ions.
引用
收藏
页码:315 / 324
页数:10
相关论文
共 12 条
[1]  
Collins Y. E., 1989, METAL IONS BACTERIA, P31
[2]  
Hughes MN, 1989, METALS MICROORGANISM
[3]  
JORGENSEN SE, 1979, STUDIES ENV SCI, V5, P190
[4]   MICROBIAL ELECTRODE BOD SENSORS [J].
KARUBE, I ;
MATSUNAGA, T ;
MITSUDA, S ;
SUZUKI, S .
BIOTECHNOLOGY AND BIOENGINEERING, 1977, 19 (10) :1535-1547
[5]  
LI F, 1993, IN PRESS BIOSENSORS
[6]   STUDY OF BOD MICROBIAL SENSORS FOR WASTE-WATER TREATMENT CONTROL [J].
LI, YR ;
CHU, J .
APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 1991, 28-9 :855-863
[7]  
SILVER S, 1989, METAL IONS BACTERIA, P121
[8]   MICROBIAL MEMBRANE-MODIFIED DISSOLVED-OXYGEN PROBE FOR RAPID BIOCHEMICAL OXYGEN-DEMAND MEASUREMENT [J].
TAN, TC ;
LI, F ;
NEOH, KG ;
LEE, YK .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 8 (02) :167-172
[9]   MEASUREMENT OF BOD BY INITIAL RATE OF RESPONSE OF A MICROBIAL SENSOR [J].
TAN, TC ;
LI, F ;
NEOH, KG .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 10 (02) :137-142
[10]  
VOGEL AI, 1961, QUANTITATIVE INORGAN, P415