CADMIUM TELLURIDE DETECTORS WITH THERMOELECTRIC COOLING

被引:38
作者
KHUSAINOV, AK
机构
[1] Riga Scientific Research Institute for Radioisotope Apparatus, Riga, 226005
关键词
D O I
10.1016/0168-9002(92)91194-E
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The results of the development of detecting units for spectrometry for X- gamma-rays based on CdTe crystals cooled by Peltier elements are presented in the given report. Cooling of CdTe detecting units with M-S-M and p-i-n structures to temperatures ranging from -30 to -40-degrees-C allows a noticeable improvement in both the energy resolution and stability. Detectors of the p-i n structure enable a resolution of 0.8 keV to be attained for an energy of 14.4 keV of 1.9) keV for an energy of 122 keV (Co-57 isotope) and of 6.9 keV for an energy of 662 keV (Cs-137 isotope) at high peak-to-valley values, respectively.
引用
收藏
页码:335 / 340
页数:6