USE OF SEQUENTIAL INJECTION TECHNIQUE AND ROBOTICS FOR THE AUTOMATION OF RHFXIII FLUOROMETRIC ACTIVITY ASSAY - CASE-STUDY

被引:21
作者
GUZMAN, M [1 ]
COMPTON, BJ [1 ]
机构
[1] ZYMOGENET INC,SEATTLE,WA 98105
关键词
D O I
10.1016/0039-9140(93)80119-C
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper evaluates the feasibility of two systems-the Sequential Injection (SI) analyzer and the Zymark Benchmate (ZB) robot-for the automation of a recombinant human Factor Thirteen (rhFXIII) fluorometric assay. The goal was to develop a routine analytical procedure suitable for the quality control lab environment. The experimental efforts focused on monitoring of the product formation for the condensation reaction between monodansyl cadaverin and dimethyl casein. The acquired kinetic data demonstrated that both systems are capable of automating the solution handling operations associated with the assay. Using a method developed with the SI system, samples containing 0-410 mug/ml of rhFXIII were analyzed, with a throughput of one sample per 8 min, and a total solution consumption of 0.8 ml. The relative standard deviation for 10 injections of 100 mug/ml rhFXIII sample was 0.91%. With the ZB robot, samples containing 0-2500 mug/ml of rhFXIII were analyzed, and the linear response was obtained for a concentration range between 0 and 1250 mug/ml of rhFXIII. The method had a sample throughput of one sample per 11 min and a total solution consumption of 6.3 ml for each analysis. Due to its commercial availability, the ZB system was preferred over the experimental SI system for the purpose of routine automated analysis of a large number or samples in the quality control lab environment.
引用
收藏
页码:1943 / 1950
页数:8
相关论文
共 8 条
[1]   NOVEL SINGLE STANDARD CALIBRATION AND DILUTION METHOD PERFORMED BY THE SEQUENTIAL INJECTION TECHNIQUE [J].
BARON, A ;
GUZMAN, M ;
RUZICKA, J ;
CHRISTIAN, GD .
ANALYST, 1992, 117 (12) :1839-1844
[2]   SEQUENTIAL INJECTION TECHNIQUE FOR AUTOMATION OF COMPLEX ANALYTICAL PROCEDURES - FLUOROMETRIC ASSAY OF FACTOR-XIII [J].
GUZMAN, M ;
POLLEMA, C ;
RUZICKA, J ;
CHRISTIAN, GD .
TALANTA, 1993, 40 (01) :81-87
[3]  
HURST JW, 1987, LABORATORY ROBOTICS
[4]   TRANSAMIDATING ENZYMES .2. CONTINUOUS FLUORESCENT METHOD SUITED FOR AUTOMATING MEASUREMENTS OF FACTOR-III IN PLASMA [J].
LORAND, L ;
LOCKRIDG.OM ;
CAMPBELL, LK ;
MYHRMAN, R ;
BRUNERLO.J .
ANALYTICAL BIOCHEMISTRY, 1971, 44 (01) :221-&
[5]  
MARSHALL GD, 1990, ANAL CHEM, V62, P1861
[6]   PRINCIPLES OF STOPPED-FLOW SEQUENTIAL INJECTION-ANALYSIS AND ITS APPLICATION TO THE KINETIC DETERMINATION OF TRACES OF A PROTEOLYTIC-ENZYME [J].
RUZICKA, J ;
GUBELI, T .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1680-1685
[7]  
Ruzicka J., 1988, FLOW INJECTION ANAL, Vsecond
[8]  
SKOOG DA, 1992, PRINCIPLES INSTRUMEN, P681