ADAPTIVE ESTIMATION OF THE MEAN FREQUENCY OF A DOPPLER SIGNAL FROM SHORT DATA WINDOWS

被引:7
作者
HERMENT, A [1 ]
DEMOMENT, G [1 ]
GUGLIELMI, JP [1 ]
DUMEE, P [1 ]
PELLOT, C [1 ]
机构
[1] UNIV PARIS SUD,ECOLE SUPER ELECT,CNRS,L2S,GIF SUR YVETTE,FRANCE
关键词
COLOR DOPPLER; MEAN FREQUENCY; CORRELATION FUNCTION;
D O I
10.1016/0301-5629(91)90085-B
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The color Doppler estimator (CE1), which is calculated from the phase of the first correlation lag of the Doppler signal, is compared to the general mean frequency estimator (CE(n)), which is based on a weighted summation of all the available correlation lags, for long and short Doppler data sets (typically 48 and 8 Doppler samples). A new estimator of the Doppler signal mean frequency is derived from the results of this study. It optimizes the compromise between the range of analyzable frequencies and the estimation variance for the characteristics of the Doppler signal. Demonstration is provided that the behavior of this estimator shifts from that of CE1 to that of CE(n), according to the setting of a single parameter. An adaptive version of this estimator is implemented and applied to Doppler recordings. Applications can be contemplated for color Doppler imaging.
引用
收藏
页码:901 / 919
页数:19
相关论文
共 23 条
[1]   DISCRETE-TIME ESTIMATION OF THE MEAN DOPPLER FREQUENCY IN ULTRASONIC BLOOD VELOCITY-MEASUREMENTS [J].
ANGELSEN, BAJ ;
KRISTOFFERSEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (04) :207-214
[2]   ZEROCROSSING DENSITY FOR ULTRASONIC DOPPLER SIGNALS OBTAINED FROM COMPUTER-SIMULATIONS [J].
ANGELSEN, BAJ ;
KRISTOFFERSEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1983, 9 (06) :661-665
[3]   INSTANTANEOUS MEASUREMENT OF BLOODFLOW BY ULTRASONIC MEANS [J].
ARTS, MGJ ;
ROEVROS, JMJ .
MEDICAL & BIOLOGICAL ENGINEERING, 1972, 10 (01) :23-&
[4]   A NEW TIME DOMAIN TECHNIQUE FOR VELOCITY-MEASUREMENTS USING DOPPLER ULTRASOUND [J].
BARBER, WD ;
EBERHARD, JW ;
KARR, SG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (03) :213-229
[5]  
BRAMANTI M, 1983, IEEE T BIOMED ENG, V3, P584
[6]  
DEMOMENT G, 1988, 4TH ASSP WORKSH SPEC
[7]   THE SQUARE ROOT F POWER-SPECTRUM CENTROID DETECTOR - SYSTEM CONSIDERATIONS, IMPLEMENTATION, AND PERFORMANCE [J].
GERZBERG, L ;
MEINDL, JD .
ULTRASONIC IMAGING, 1980, 2 (03) :262-289
[8]   POWER-SPECTRUM CENTROID DETECTION FOR DOPPLER SYSTEMS APPLICATIONS [J].
GERZBERG, L ;
MEINDL, JD .
ULTRASONIC IMAGING, 1980, 2 (03) :232-261
[10]   REAL-TIME TWO-DIMENSIONAL BLOOD-FLOW IMAGING USING AN AUTO-CORRELATION TECHNIQUE [J].
KASAI, C ;
NAMEKAWA, K ;
KOYANO, A ;
OMOTO, R .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1985, 32 (03) :458-464