ULTRASOUND DOPPLER TECHNIQUE FOR THE NON-INVASIVE DETERMINATION OF THE PRESSURE-GRADIENT IN THE BJORK-SHILEY MITRAL-VALVE

被引:65
作者
HOLEN, J
SIMONSEN, S
FROYSAKER, T
机构
[1] UNIV HOSP OSLO,RIKSHOSP,SURG DEPT A,OSLO,NORWAY
[2] UNIV HOSP OSLO,RIKSHOSP,DEPT MED B,OSLO,NORWAY
关键词
D O I
10.1161/01.CIR.59.3.436
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The accuracy in determining the pressure gradient in the Bjork-Shiley mitral implant from noninvasive ultrasound Doppler data was explored in nine adult patients. Manometric and ultrasound data were collected simultaneously, and identical diastolic periods were used to compare the manometric gradient (Δ(M)) with the gradient obtained from ultrasound data (Δ P(U)). In the nine patients the mean diastolic value of ΔP(M) ranged from 2-12.5 mm Hg and the difference between the mean diastolic values of ΔP(M) and ΔP(U) was 0.3 ± 1.0 mm Hg(SD). The results of the investigation indicated that the method is accurate and reliable in the noninvasive determination of the mean diastolic gradient in the Bjork-Shiley mitral implant.