STUDY OF THE SPATIAL-RESOLUTION IN GAMMA-DETECTORS USING PHOTOSENSITIVE WIRE CHAMBERS IN VIEW OF THEIR USE AS HIGH-RESOLUTION PET SCANNERS

被引:5
作者
BRUYNDONCKX, P
SHUPING, Z
TAVERNIER, S
机构
[1] Interuniversity Institute for High Energies, VUB-ULB, 1050 Brussels
关键词
D O I
10.1016/0168-9002(92)90268-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The spatial resolution in gamma detectors using BaF2 scintillators and photosensitive wire chambers was studied for 511 keV gamma rays with different crystal geometries. When a matrix of small crystals was used, the spatial resolution was dominated by the crystal dimensions. With crystals measuring 3 x 3 x 20 and 2 x 2 x 15 mm3 and with 0.3 mm dead space between them, we obtained a resolution FWHM of 3.7 and 3.0 mm, respectively. With one large crystal 20 mm thick a resolution of 7.9 mm FWHM was obtained. We used a Monte Carlo study to evaluate the performances of a high resolution PET scanner for small animals based on this principle. Such a device is presently under construction.
引用
收藏
页码:54 / 59
页数:6
相关论文
共 10 条
[1]   COUPLING OF A BAF2 SCINTILLATOR TO A TMAE PHOTO-CATHODE AND A LOW-PRESSURE WIRE CHAMBER [J].
ANDERSON, DF ;
BOUCLIER, R ;
CHARPAK, G ;
MAJEWSKI, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 217 (1-2) :217-223
[2]   A RING IMAGING CHERENKOV DETECTOR, THE DELPHI BARREL RICH PROTOTYPE .B. EXPERIMENTAL STUDIES OF THE DETECTOR PERFORMANCE FOR PARTICLE IDENTIFICATION [J].
ARNOLD, R ;
BAILLON, P ;
BESCH, HJ ;
BOSTEELS, M ;
CHRISTOPHEL, E ;
DRACOS, M ;
GIOMATARIS, Y ;
GUYONNET, JL ;
PASSARDI, G ;
PETROFF, P ;
SEGUINOT, J ;
YPSILANTIS, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 270 (2-3) :289-318
[3]   BAF2 SCINTILLATORS WITH WIRE CHAMBER READOUT FOR POSITRON EMISSION TOMOGRAPHY [J].
BRUYNDONCKX, P ;
DEBRUYNE, J ;
ETIENNE, L ;
GRUWE, M ;
GUERARD, B ;
TAVERNIER, S ;
SHUPING, Z .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :107-115
[4]   USE OF TMAE IN A MULTISTEP PROPORTIONAL CHAMBER FOR CHERENKOV RING IMAGING AND OTHER APPLICATIONS [J].
CHARPAK, G ;
SAULI, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 225 (03) :627-635
[6]  
DERENZO SE, 1979, POSITRON ANNIHILATIO, P819
[7]   LOW-PRESSURE ULTRAVIOLET PHOTON DETECTOR WITH TMAE GAS PHOTOCATHODE [J].
MAJEWSKI, S ;
ANDERSON, DF ;
CONSTANTAFANOURAKIS, P ;
KROSS, B ;
FANOURAKIS, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 264 (2-3) :235-250
[8]   EVALUATION OF SYSTEMATIC-ERRORS IN THE AVALANCHE LOCALIZATION ALONG THE WIRE WITH CATHODE STRIPS READOUT MWPC [J].
PIUZ, F ;
ROOSEN, R ;
TIMMERMANS, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 196 (2-3) :451-462
[9]   CENTROID FINDING METHOD FOR POSITION-SENSITIVE DETECTORS [J].
RADEKA, V ;
BOIE, RA .
NUCLEAR INSTRUMENTS & METHODS, 1980, 178 (2-3) :543-554
[10]  
TAVERNIER S, 1992, PHYS MED BIOL, V37, P625