HEAT AND MOISTURE EXCHANGERS AND VAPORIZING HUMIDIFIERS IN THE INTENSIVE-CARE UNIT

被引:130
作者
MARTIN, C [1 ]
PERRIN, G [1 ]
GEVAUDAN, MJ [1 ]
SAUX, P [1 ]
GOUIN, F [1 ]
机构
[1] HOP SAN SALVADOR, DEPT MICROBIOL, MARSEILLE, FRANCE
关键词
D O I
10.1378/chest.97.1.144
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
A prospective, randomized, controlled study was undertaken to compare the Pall Ultipor breathing circuit filter (PUBCF), a heat-and-moisture exchanger, and heated hot water systems (HHWSs) in ICU patients submitted to controlled mechanical ventilation. Humidification of inspired gas and bacterial contamination of breathing circuits were evaluated. During the study, there were six episodes of tracheostomy tube (TT) occlusion in six patients included in the PUBCF group. No patient out of 42 included in the HHWS group experienced this complication (p < 0.01). There were 4 percent of days with thick and tenacious bronchial secretions in the PUBCF group and no case in the HHWS group (p < 0.02). In the PUBCF group, 23 percent of days with hypothermia were noted as opposed to 12 percent in the HHWS group (p < 0.01). Fewer breathing circuits were found to be contaminated in the PUBCF group (11 percent) than in the HHWS group (54 percent, p < 0.01). In patients with an organism growing in bronchial specimens, the same organism was found to contaminate the breathing circuit in 10 percent of cases in the PUBCF group and 77 percent of cases in the HHWS (p < 0.01). We conclude that, in the conditions of this study, the PUBCF did not provide sufficient humidification of inspired gas in ICU patients. Protection against contamination of breathing circuits was effective, but 10 percent of patients remained at risk for this complication.
引用
收藏
页码:144 / 149
页数:6
相关论文
共 33 条
[1]   MOISTURE EXCHANGERS DO NOT PREVENT PATIENT CONTAMINATION OF VENTILATORS - A MICROBIOLOGICAL STUDY [J].
BYGDEMAN, S ;
VONEULER, C ;
NYSTROM, B .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1984, 28 (06) :591-594
[2]   EFFECTS OF DRY ANESTHETIC GASES ON TRACHEOBRONCHIAL CILIATED EPITHELIUM [J].
CHALON, J ;
MALEBRANCHE, J ;
LOEW, DAY .
ANESTHESIOLOGY, 1972, 37 (03) :338-+
[3]   HUMIDITY AND THE ANESTHETIZED PATIENT [J].
CHALON, J ;
PATEL, C ;
ALI, M ;
RAMANATHAN, S ;
CAPAN, L ;
TANG, CK ;
TURNDORF, H .
ANESTHESIOLOGY, 1979, 50 (03) :195-198
[4]  
CHALON J, 1984, ANESTH ANALG, V63, P566
[5]  
CHALON J, 1981, HUMIDIFICATION ANEST, P16
[6]   ENDOTRACHEAL-TUBE OCCLUSION ASSOCIATED WITH THE USE OF HEAT AND MOISTURE EXCHANGERS IN THE INTENSIVE-CARE UNIT [J].
COHEN, IL ;
WEINBERG, PF ;
FEIN, IA ;
ROWINSKI, GS .
CRITICAL CARE MEDICINE, 1988, 16 (03) :277-279
[7]   CONTAMINATION OF MECHANICAL VENTILATORS WITH TUBING CHANGES EVERY 24 OR 48 HOURS [J].
CRAVEN, DE ;
CONNOLLY, MG ;
LICHTENBERG, DA ;
PRIMEAU, PJ ;
MCCABE, WR .
NEW ENGLAND JOURNAL OF MEDICINE, 1982, 306 (25) :1505-1509
[8]  
CRAVEN DE, 1984, AM REV RESPIR DIS, V129, P625
[9]   EVOLUTION OF HEAT AND MOISTURE IN RESPIRATORY TRACT DURING ANESTHESIA WITH A NON-REBREATHING SYSTEM [J].
DERY, R .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1973, 20 (03) :296-309
[10]  
FLOURNOY DJ, 1980, RESP CARE, V25, P742