TRANSCUTANEOUS PCO2 AND PO2 - A MULTICENTER STUDY OF ACCURACY

被引:63
作者
PALMISANO, BW [1 ]
SEVERINGHAUS, JW [1 ]
机构
[1] UNIV CALIF SAN FRANCISCO,1386 HSE,SAN FRANCISCO,CA 94143
来源
JOURNAL OF CLINICAL MONITORING | 1990年 / 6卷 / 03期
关键词
Blood gas analysis; Measurement techniques: blood gases; Monitoring;
D O I
10.1007/BF02832145
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A multicentcr study used 756 samples from 251 patients in 12 institutions to compare arterial (PaO2, PaCO2) with transcutaneous (PsO2, PsCO2) oxygen and carbon dioxide tensions, measured usually at 44°C. Of these samples, 336 were obtained from 116 neonates, 27 from 25 children with cystic fibrosis, and 140 from 40 patients under general anesthesia. Ninety-one patients were between 4 weeks and 18 years of age, 32 were between 18 and 60 years, and 12 were over 60. The ratio of transcutaneous to arterial P(s/a)CO2 was 1.01 ±0.11 with PaCO2 less than 30 mm Hg, increasing to 1.04 ±0.08 at PaCO2 greater than 40 mm Hg. Mean bias and its standard deviation (PsCO2 - PaCO2) were + 1.3 ± 3.9 mm Hg in the entire group, + 1.8 ± 4.2 mm Hg in neonates (NS). Bias was +0.2 ± 2.7 mm Hg when PaCO2 was less than 30 mm Hg (N = 175, NS), 1.0 ± 3.4 with 30 < PaCO2 < 40 (n = 329, p < 0.001), and +2.04 ± 4.00 mm Hg with 40 < PaCO2 < 70 (n = 229, p < 0.001). These data suggest that, using transcutaneous PCO2 monitors with inbuilt temperature correction of 4.5%/‡C, the skin metabolic offset should be set to 6 mm Hg. The linear regression was PsCO2 =1.052(PaCO2)-0.56, Sy·x = 3.92, R = 0.929 (n = 756); and PsCO2 = 1.09(PaCO2)-1.57, Sy·x = 4.17, R = 0.928 in neonates (n = 336). The use of vasopressors and vasodilators had no significant effect on bias or its standard deviation or on regression slope and intercept (n = 78). In cystic fibrosis patients, bias and standard deviation were 0.0 ± 1.7 mm Hg (n = 27). Under anesthesia, PsCO2 = 1.07PaCO2-1.58, with bias and standard deviation = 0.6 ± 3.5 (n = 140). For oxygen, at PaO2 ≤ 80 the ratio P(s/a)O2 = 1.05 ± 0.16 in nconates and 0.93 ± 0.21 in older patients, but when PaO2 > 80, P(s/a)O2 fell to 0.88 ± 0.18 in neonates and 0.74 ± 0.21 in older patients. The errors were significantly greater (p < 0.001) in older patients than in neonates above but not below 80 mm Hg, and within both groups errors were significantly greater above than below 80 mm Hg. © 1990 Little, Brown and Company.
引用
收藏
页码:189 / 195
页数:7
相关论文
共 15 条
[1]  
Tremper K.K., Transcutaneous PO<sub>2</sub> measurement, Can Anaesth Soc J, 31, pp. 664-677, (1984)
[2]  
Tremper K.K., Barker S.J., Transcutaneous oxygen measurement: experimental studies and adult applications, Int Anesthesiol Clin, 25, 3, pp. 67-96, (1987)
[3]  
Phan C.Q., Tremper K.K., Lee S.E., Barker S.J., Noninvasive monitoring of carbon dioxide: a comparison of the partial pressure of transcutaneous and end-tidal carbon dioxide with the partial pressure of arterial carbon dioxide, J Clin Monit, 3, pp. 149-154, (1987)
[4]  
McEvedy B.A.B., McLeod M.E., Mulera M., Et al., End-tidal, transcutaneous and arterial PCO<sub>2</sub> measurements in critically ill neonates: a comparative study, Anethesiology, 69, pp. 112-116, (1988)
[5]  
Severinghaus J.W., A combined transcutaneous PO<sub>2</sub>-PCO<sub>2</sub> electrode with electrochemical HCO<sub>3</sub> stabilization, J Appl Physiol, 51, pp. 1027-1032, (1981)
[6]  
Thunstrom A.M., Stafford M.J., Severinghaus J.W., A two temperature, two PO<sub>2</sub> method of estimating the determinants of tcPO<sub>2</sub>, Continuous transcutaneous blood gas monitoring. Birth Defects, pp. 167-182, (1979)
[7]  
Hazinski T.A., Severinghaus J.W., Transcutaneous analysis of arterial PCO<sub>2</sub>, Med Instr, 16, pp. 150-153, (1982)
[8]  
Glenski J.A., Cucchiara R.F., Transcutaneous O<sub>2</sub> and CO<sub>2</sub> monitoring of ncurosurgical patients: detection of air embolism, Anesthesiology, 64, pp. 546-550, (1986)
[9]  
Mindt W., Eberhard P., SchAfer R., Methodologic factors affecting the relationship between transcutaneous and arterial PCO<sub>2</sub>, Continuous transcutaneous blood gas monitoring, pp. 199-215, (1983)
[10]  
Cassady G., Transcutaneous monitoring in the newborn infant, J Pediatr, 103, pp. 837-848, (1983)