EFFECTS OF PACING RATE AND TIMING OF DEFIBRILLATION SHOCK ON THE RELATION BETWEEN THE DEFIBRILLATION THRESHOLD AND THE UPPER LIMIT OF VULNERABILITY IN OPEN CHEST DOGS

被引:69
作者
CHEN, PS
FELD, GK
MOWER, MM
PETERS, BB
机构
[1] VET AFFAIRS MED CTR,SAN DIEGO,CA
[2] AMER HEART ASSOC,DALLAS,TX
关键词
D O I
10.1016/0735-1097(91)90690-B
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the relation between the defibrillation threshold and the upper limit of vulnerability, the shock strength associated with 50% probability of successful defibrillation (DFT50) and that associated with 50% probability of reaching the upper limit of vulnerability (ULV50) were determined in 20 open chest dogs with use of the delayed up-down method, with pacing drive cycle lengths of 150 to 500 ms and either single 6-ms shocks (10 dogs) or 12-ms biphasic shocks (10 dogs) given at the mid-upslope, peak and mid-downslope of the T wave of electrocardiographic lead II. The shocks were given by means of a patch-patch configuration on the anterior and posterior surfaces of the heart, which was paced from a stimulating electrode attached to the left ventricular apex. Analysis of variance showed no statistically significant differences in ULV50 as determined with different pacing cycle lengths. For monophasic shocks, DFT50 (331 +/- 66 V or 5.8 +/- 2.7 J) was not significantly different from ULV50 determined at the mid-upslope of the T wave (318 +/- 64 V or 5 +/- 2 J). The correlation coefficients between the two values were 0.74 (p = 0.014) for voltage and 0.67 (p = 0.034) for energy. In contrast, DFT50 was significantly higher than ULV50 as determined at the peak of the T wave (219 +/- 43 V or 2.3 +/- 1 J) and mid-downslope of the T wave (200 +/- 38 V or 1.9 +/- 0.9 J). In three dogs, ventricular fibrillation could not be induced at the mid-downslope of the T wave with any baseline pacing (S1) cycle length. For biphasic shocks, DFT50 (285 +/- 53 V or 4 +/- 1.5 J) was not significantly different from ULV50 as determined at the mid-upslope of the T wave (293 +/- 54 V or 3.9 +/- 1.2 J). The correlation coefficients between the two values were 0.90 (p < 0.001) for voltage and 0.82 (p = 0.004) for energy. In contrast, DFT50 Was significantly higher than ULV50 determined at the peak of the T wave (208 +/- 31 V or 1.9 +/- 0.5 J) and mid-downslope of the T wave (180 +/- 16 V or 1.7 +/- 0.8 J). In six dogs, ventricular fibrillation could not be induced at the mid-downslope of the T wave with any S1 cycle length. It is concluded that with either monophasic or biphasic shocks, DFT50 can be accurately estimated by ULV50 as determined at the mid-upslope of the T wave. Altering the baseline pacing cycle length does not affect this relation.
引用
收藏
页码:1555 / 1563
页数:9
相关论文
共 25 条
  • [1] EFFECTS OF ACTIVATION SEQUENCE ON LOCAL RECOVERY OF VENTRICULAR EXCITABILITY IN DOG
    ABILDSKOV, JA
    [J]. CIRCULATION RESEARCH, 1976, 38 (04) : 240 - 243
  • [2] AMLIE JP, 1979, ACTA PHARMACOL TOX, V44, P264
  • [3] AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR DISCHARGES AND ACUTE MYOCARDIAL INJURY
    AVITALL, B
    PORT, S
    GAL, R
    MCKINNIE, J
    TCHOU, P
    JAZAYERI, M
    TROUP, P
    AKHTAR, M
    [J]. CIRCULATION, 1990, 81 (05) : 1482 - 1487
  • [4] EFFECT OF PENTOBARBITAL ANESTHESIA ON VENTRICULAR DEFIBRILLATION THRESHOLD IN DOGS
    BABBS, CF
    [J]. AMERICAN HEART JOURNAL, 1978, 95 (03) : 331 - 337
  • [5] EPICARDIAL ACTIVATION AND REPOLARIZATION PATTERNS IN PATIENTS WITH RIGHT VENTRICULAR HYPERTROPHY
    CHEN, PS
    MOSER, KM
    DEMBITSKY, WP
    AUGER, WR
    DAILY, PO
    CALISI, CM
    JAMIESON, SW
    FELD, GK
    [J]. CIRCULATION, 1991, 83 (01) : 104 - 118
  • [6] COMPARISON OF THE DEFIBRILLATION THRESHOLD AND THE UPPER LIMIT OF VENTRICULAR VULNERABILITY
    CHEN, PS
    SHIBATA, N
    DIXON, EG
    MARTIN, RO
    IDEKER, RE
    [J]. CIRCULATION, 1986, 73 (05) : 1022 - 1028
  • [7] THE RELATIONSHIP BETWEEN SUCCESSFUL DEFIBRILLATION AND DELIVERED ENERGY IN OPEN-CHEST DOGS - REAPPRAISAL OF THE DEFIBRILLATION THRESHOLD CONCEPT
    DAVY, JM
    FAIN, ES
    DORIAN, P
    WINKLE, RA
    [J]. AMERICAN HEART JOURNAL, 1987, 113 (01) : 77 - 84
  • [8] IMPROVED DEFIBRILLATION THRESHOLDS WITH LARGE CONTOURED EPICARDIAL ELECTRODES AND BIPHASIC WAVE-FORMS
    DIXON, EG
    TANG, ASL
    WOLF, PD
    MEADOR, JT
    FINE, MJ
    CALFEE, RV
    IDEKER, RE
    [J]. CIRCULATION, 1987, 76 (05) : 1176 - 1184
  • [9] FABIATO PA, 1967, ARCH MAL COEUR VAISS, V60, P527
  • [10] KAVANAGH KM, 1990, PACE, V13, P536