CALCULATION OF PULSE-WAVE VELOCITY USING CROSS-CORRELATION - EFFECTS OF REFLEXES IN THE ARTERIAL TREE

被引:94
作者
BENTHIN, M
DAHL, P
RUZICKA, R
LINDSTROM, K
机构
[1] Department of Electrical Measurements, Lund Institute of Technology, S-221 00 Lund
关键词
ULTRASOUND; PULSE-WAVE VELOCITY (PWV); CROSS CORRELATION; REFLECTED PULSE WAVES;
D O I
10.1016/0301-5629(91)90182-V
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The need to investigate the elastic properties of arterial vessels in situ for the early recognition of degenerative disorders in arteries has long been apparent. Pulse-wave diagnostics has been used in traditional medicine, but only in a qualitative way. By measuring the vessel diameter change, simultaneously at two levels along an artery, it should be possible to calculate the pulse-wave velocity (PWV). A newly developed ultrasonic instrument, DIAMOVE, was designed to provide a two-dimensional, real time, B-mode image for display of the exact position of two measurement sites; a computer-controlled digital phased-locked loop detector for the measurement of diameter changes; and a personal computer for the display of the momentary displacement of the artery walls. The cross-correlation technique was used on the information from the vessel wall movements to calculate the resulting PWV. Early measurements on healthy patients showed some unexpected artifacts, which were found to be due to reflected waves, affecting the shape of the pulse-wave at the sites of measurement. A modified PWV estimation algorithm, where only the foot of the pulse-wave in every beat is utilized, seems to have solved this problem and now makes the method suitable for clinical use.
引用
收藏
页码:461 / 469
页数:9
相关论文
共 25 条
  • [1] DIAMETER OF INTACT CAROTID ARTERY IN MAN AND ITS CHANGE WITH PULSE PRESSURE
    ARNDT, JO
    KLAUSKE, J
    MERSCH, F
    [J]. PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1968, 301 (03): : 230 - &
  • [2] ASBERG A, 1967, ULTRASONICS 0405, P113
  • [3] BEDFORD DE, 1951, BRIT HEART J, V13, P423
  • [4] BUSCHMANN W, 1964, OPHTALMOLOGIE, V166, P519
  • [5] EDLER I, 1971, CLIN APPLICATIONS UL, V3, P455
  • [6] Edler I, 1954, KUNGLIGA FYSIOGRAFIS, V24, P5
  • [7] EIKNES SH, 1980, THESIS U MALMO, P162
  • [8] EISENBERG D, 1987, QI MOTE MED KINESISK, P42
  • [9] PULSE-WAVE RECORDING - DEVELOPMENT OF A METHOD FOR INVESTIGATING FETAL CIRCULATION INUTERO
    ERIKSEN, PS
    GENNSER, G
    LINDSTROM, K
    BENTHIN, M
    DAHL, P
    [J]. JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1985, 9 (01) : 18 - 27
  • [10] PHYSIOLOGICAL-CHARACTERISTICS OF DIAMETER PULSES IN THE FETAL DESCENDING AORTA
    ERIKSEN, PS
    GENNSER, G
    LINDSTROM, K
    [J]. ACTA OBSTETRICIA ET GYNECOLOGICA SCANDINAVICA, 1984, 63 (04) : 355 - 363