INFLUENCE OF ERYTHROCYTE VELOCITY ON IMPEDANCE PLETHYSMOGRAPHIC MEASUREMENTS

被引:45
作者
PEURA, RA [1 ]
PENNEY, BC [1 ]
ARCURI, J [1 ]
ANDERSON, FA [1 ]
WHEELER, HB [1 ]
机构
[1] UNIV MASSACHUSETTS,SCH MED,DEPT SURG,WORCESTER,MA 01604
关键词
Blood impedance; Impedance plethysmography; Rheography;
D O I
10.1007/BF02451914
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Impedance plethysmography has been widely used clinically. However, uncertainty exists concerning the source of the impedance signal. The present study was designed to indicate the relative contributions made by blood volume changes and blood flow changes. An in vitro flow system containing a pulsatile pump, a bovine carotid artery, a rigid tube, and a terminal resis tance was assembled. Impedance recordings were made on both the expansile artery and the rigid tube. The impedance signal from the rigid tube was taken as the flow-related component. The signal from the expansile artery represented the combined effects of volume changes and flow changes. The recordings were made over the full physiological range of pulse pressures and peak flow values, using both normal saline and blood (haematocrit =26±4%). No impedance change was detected in the rigid vessel using saline as the perfusate. When blood was used, the impedance signal from the rigid tube averaged about 10% of that from the expansile vessel and its morphology was different. Changes in erthrocyte velocity produced a small but definite change in the impedance signal recorded from the artery. © 1978 IFMBE.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 18 条
[1]  
ANDERSON FA, 1977, 12TH P ASS ADV MED I
[2]   CONDUCTIVITY OF SHEARED SUSPENSIONS OF ELLIPSOIDAL PARTICLES WITH APPLICATION TO BLOOD-FLOW [J].
EDGERTON, RH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1974, BM21 (01) :33-43
[3]   EFFECT OF FREQUENCY CHANGES ON ELECTRICAL CONDUCTANCE OF MOVING AND STATIONARY BLOOD [J].
FREWER, RA .
MEDICAL & BIOLOGICAL ENGINEERING, 1972, 10 (06) :734-741
[4]  
FRICKE HUGO, 1926, JOUR GEN PHYSIOL, V9, P153
[5]   SPECIFIC RESISTANCE OF CANINE BLOOD AT BODY-TEMPERATURE [J].
GEDDES, LA ;
DACOSTA, CP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (01) :51-53
[6]   ELECTRICAL IMPEDANCE OF PULSATILE BLOOD FLOW IN RIGID TUBES AND IN ISOLATED ORGANS [J].
GOLLAN, F ;
NAMON, R .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1970, 170 (02) :568-&
[7]  
HIRSH EM, 1974, FED PROC, V33, P334
[8]  
KANAI H, 1973, 10TH DIG INT C MED B
[9]  
NYBOER J, 1959, ELECTRICAL IMPEDANCE
[10]  
POWERS SR, 1958, SURGERY, V44, P53