DETECTION OF CHANGES IN ATRIAL ENDOCARDIAL ACTIVATION WITH USE OF AN ORTHOGONAL CATHETER

被引:13
作者
GERSTENFELD, EP
SAHAKIAN, AV
BAERMAN, JM
ROPELLA, KM
SWIRYN, S
机构
[1] EVANSTON HOSP CORP,DEPT MED,DIV CARDIOL,ROOM 300 BURCH HALL,2650 RIDGE AVE,EVANSTON,IL 60201
[2] NORTHWESTERN UNIV,DEPT BIOMED ENGN,EVANSTON,IL 60201
[3] NORTHWESTERN UNIV,SCH MED,FEINBERG CARDIOVASC RES INST,CHICAGO,IL 60611
关键词
D O I
10.1016/0735-1097(91)90763-Y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The ability of a catheter with an orthogonal electrode configuration to sense differences in the direction of local atrial endocardial activation was tested in 18 consecutive patients with intact retrograde conduction. In all 18, discrimination of anterograde from retrograde conduction at a single atrial site was examined; in 5 of the 18, multiple sites were examined to determine if the discriminatory ability of the catheter was site dependent. The catheter was specially designed with bipoles in the x, y and z directions. A vector was computed for each electrogram during anterograde and retrograde conduction. Electrogram amplitude along the standard bipole was also compared for anterograde and retrograde conduction. Mean electrogram amplitude for the standard bipole was significantly different for anterograde than for retrograde conduction in 17 of 18 patients (mean +/- SD 4 +/- 1.9 vs. 2.7 +/- 1.3 mV; p < 0.005), with complete separation of amplitude distributions in 4 patients. The electrogram vector during anterograde conduction was significantly different from that during retrograde conduction in all 18 patients (p < 0.0001), with complete separation of vector distributions in 14. In some patients with multiple site recordings, the choice of site greatly affected separation based on electrogram amplitude or vector, or both. The orthogonal catheter can be used to sense directional differences in local endocardial activation. The catheter shows promise for discriminating anterograde from retrograde conduction and examining the direction of endocardial activation in the heart during an electrophysiologic examination.
引用
收藏
页码:1034 / 1042
页数:9
相关论文
共 24 条
[1]  
AKHTAR M, 1985, CLIN PROG PACING ELE, V3, P90
[2]   EFFECT OF BIPOLE CONFIGURATION ON ATRIAL ELECTROGRAMS DURING ATRIAL-FIBRILLATION [J].
BAERMAN, JM ;
ROPELLA, KM ;
SAHAKIAN, AV ;
KIRSH, JA ;
SWIRYN, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (01) :78-87
[3]  
BLANCHARD SM, 1990, WAVEFRONT ORIENTATIO, P111
[4]   TOWARD OPTIMIZING THE DETECTION OF ATRIAL DEPOLARIZATION WITH FLOATING BIPOLAR ELECTRODES [J].
BROWNLEE, RR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (03) :431-442
[5]   TOTAL EXCITATION OF ISOLATED HUMAN HEART [J].
DURRER, D ;
VANDAM, RT ;
FREUD, GE ;
JANSE, MJ ;
MEIJLER, FL ;
ARZBAECHER, RC .
CIRCULATION, 1970, 41 (06) :899-+
[6]   AN ACCURATE, CLINICALLY PRACTICAL SYSTEM FOR SPATIAL VECTORCARDIOGRAPHY [J].
FRANK, E .
CIRCULATION, 1956, 13 (05) :737-749
[7]   ENDLESS LOOP TACHYCARDIA IN AN AV UNIVERSAL (DDD) PACEMAKER [J].
FURMAN, S ;
FISHER, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (04) :486-489
[8]   A NEW ORTHOGONAL LEAD FOR P-SYNCHRONOUS PACING [J].
GOLDREYER, BN ;
OLIVE, AL ;
LESLIE, J ;
CANNOM, DS ;
WYMAN, MG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (06) :638-644
[9]   ORTHOGONAL ELECTROGRAM SENSING [J].
GOLDREYER, BN ;
KNUDSON, M ;
CANNOM, DS ;
WYMAN, MG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02) :464-469
[10]   ACTIVATION PATTERNS IN HEALED EXPERIMENTAL MYOCARDIAL-INFARCTION [J].
KADISH, A ;
BALKE, CW ;
LEVINE, JF ;
MOORE, EN ;
SPEAR, JF .
CIRCULATION RESEARCH, 1989, 65 (06) :1698-1709