MICROSEQUENCING OF PROTEINS AND PEPTIDES IN THE KNAUER SEQUENCER WITH AND WITHOUT COVALENT ATTACHMENT TO POLYVINYLIDENE DIFLUORIDE MEMBRANES BY THE WET-PHASE DEGRADATION TECHNIQUE

被引:12
作者
HERFURTH, E [1 ]
PILLING, U [1 ]
WITTMANLIEBOLD, B [1 ]
机构
[1] MAX PLANCK INST MOLEC GENET, ABTEILUNG WITTMANN, IHNESTR 73, W-1000 BERLIN 33, GERMANY
来源
BIOLOGICAL CHEMISTRY HOPPE-SEYLER | 1991年 / 372卷 / 05期
关键词
PROTEIN AND PEPTIDE SEQUENCING; WET-PHASE EDMAN DEGRADATION; SOLID-PHASE MICRO SEQUENCE ANALYSIS; COVALENT ATTACHMENT TO DITC-PVDF MEMBRANES; KNAUER MODULAR SEQUENCER; CROSS-FLOW-REACTOR;
D O I
10.1515/bchm3.1991.372.1.351
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Proteins and large peptides were degraded with phenylisothiocyanate (PITC) in the horizontal flow-through-reactor of the Modular Knauer Sequencer (Fischer, S., Reimann, F. & Wittmann-Liebold, B. (1989) in Methods in Protein Sequence Analysis (Wittmann-Liebold, B., ed.) Springer-Verlag, Berlin, pp. 98-107) by the wet-phase filter technique (Wittmann-Liebold, B. (1988) J. Prot. Chem. 7, 224-225) employing polyvinylidene difluoride (PVDF) membranes without polybrene. In order to prevent losses of small peptides during solvent washes at the degradation, 1,4-phenylene diisothiocyanate (DITC) derivatized PVDF support (MilliGen, Burlington, MA) was used to covalently attach the peptide via its lysine groups in situ within the cross-flow reaction chamber onto this membrane (Herfurth, E., Pilling, U. & Wittmann-Liebold, B. (1990) J. Prot. Chem. 9, 267). We found these membranes very suitable for peptide degradations in the Knauer sequencer. In almost all cases we were able to identify the amino-acid residues of the peptide up to its last covalent fixation point to the membrane.
引用
收藏
页码:351 / 361
页数:11
相关论文
共 31 条
  • [1] AEBERSOLD RH, 1986, J BIOL CHEM, V261, P4229
  • [2] ASHMAN K, 1985, MODERN METHODS ANAL, P151
  • [3] ASHMAN K, 1986, ADV METHODS PROTEIN, P219
  • [4] PROTEIN BLOTTING FOLLOWED BY MICROSEQUENCING
    CHOLI, T
    WITTMANNLIEBOLD, B
    [J]. ELECTROPHORESIS, 1990, 11 (07) : 562 - 568
  • [5] BLOTTING OF PROTEINS ONTO IMMOBILON MEMBRANES - INSITU CHARACTERIZATION AND COMPARISON WITH HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    CHOLI, T
    KAPP, U
    WITTMANNLIEBOLD, B
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 476 : 59 - 72
  • [6] Coull JM, 1989, METHODS PROTEIN SEQU, P69
  • [7] ECKART K, 1988, J PROTEIN CHEM, V7, P221
  • [8] METHOD FOR DETERMINATION OF THE AMINO ACID SEQUENCE IN PEPTIDES
    EDMAN, P
    [J]. ACTA CHEMICA SCANDINAVICA, 1950, 4 (02): : 283 - 293
  • [9] A PROTEIN SEQUENATOR
    EDMAN, P
    BEGG, G
    [J]. EUROPEAN JOURNAL OF BIOCHEMISTRY, 1967, 1 (01): : 80 - &
  • [10] FISCHER S, 1989, METHODS PROTEIN SEQU, P98