8 YEARS EXPERIENCE WITH THE MEDTRONIC-HALL VALVE PROSTHESIS

被引:41
作者
VALLEJO, JL
GONZALEZSANTOS, JM
ALBERTOS, J
RIESGO, MJ
BASTIDA, ME
RICO, MJ
GONZALEZDIEGO, F
ARCAS, R
机构
[1] UNIV COMPLUTENSE MADRID, GEN HOSP GREGORIO MARANON, SCH MED, DEPT CARDIAC SURG, MADRID 3, SPAIN
[2] UNIV COMPLUTENSE MADRID, GEN HOSP GREGORIO MARANON, SCH MED, DEPT ANAESTHESIA, MADRID 3, SPAIN
关键词
D O I
10.1016/0003-4975(90)90489-S
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
During the period January 1981 to September 1986, 444 Medtronic-Hall heart valve prostheses were implanted in 351 patients (mean age, 45 ± 10 years) mainly for rheumatic valve disease (63.2%). Most of the patients were in New York Heart Association functional class III. Concomitant surgical procedures, mainly conservative tricuspid or mitral procedures or coronary artery bypass grafting, were performed in 101 patients (28.7%). Single-valve replacement was performed in 262 patients (74.6%) (aortic in 117 patients, mitral in 143, and tricuspid in 2), double-valve replacement in 85 (24.2%) (mitral and aortic in 83 and mitral and tricuspid in 2), and triple-valve replacement in 4 (1.1%). Hospital mortality was 6.2%. Follow-up was 97.7% complete. The overall actuarial 8-year survival rate was 77.2%. The linearized incidence of valve-related complications was as follows: thromboembolism, 1.5%/patient-year; reoperation, 1.5%/ patient-year; endocarditis, 1.25%/patient-year; hemolysis, 0.52%/patient-year; anticoagulant-related hemorrhage, 0.39%/patient-year; and noninfection-related paraprosthetic leak, 0.33%/patient-year. There were no instances of structural failure. We conclude that after 8 years of follow-up, the Medtronic-Hall valve prosthesis has an excellent clinical performance and a low range of valve-related complications. © 1990.
引用
收藏
页码:429 / 436
页数:8
相关论文
共 30 条
  • [1] ANTUNES MJ, 1988, J THORAC CARDIOV SUR, V95, P980
  • [2] ARIS A, 1988, J THORAC CARDIOV SUR, V95, P423
  • [3] ST JUDE MEDICAL PROSTHESIS - VALVE-RELATED DEATHS AND COMPLICATIONS
    AROM, KV
    NICOLOFF, DM
    KERSTEN, TE
    LINDSAY, WG
    NORTHRUP, WF
    [J]. ANNALS OF THORACIC SURGERY, 1987, 43 (06) : 591 - 598
  • [4] THE MEDTRONIC-HALL CARDIAC-VALVE - 7-1/2 YEARS CLINICAL-EXPERIENCE
    BEAUDET, RL
    POIRIER, NL
    DOYLE, D
    NAKHLE, G
    GAUVIN, C
    [J]. ANNALS OF THORACIC SURGERY, 1986, 42 (06) : 644 - 650
  • [5] BEAUDET RL, 1983, CONCEPTS CONTROVERSI, P134
  • [6] BJORK VO, 1979, J THORAC CARDIOV SUR, V78, P331
  • [7] 10-YEAR ANALYSIS OF THE BJORK-SHILEY STANDARD AORTIC-VALVE
    BORKON, AM
    SOULE, L
    BAUGHMAN, KL
    AOUN, H
    GARDNER, TJ
    WATKINS, L
    BAUMGARTNER, WA
    GOTT, VL
    REITZ, BA
    [J]. ANNALS OF THORACIC SURGERY, 1987, 43 (01) : 39 - 51
  • [8] BRODATY D, 1987, ARCH MAL COEUR VAISS, V80, P1719
  • [9] BUTCHART EG, 1988, CIRCULATION, V78, P66