VENTRICULAR-TACHYCARDIA AND FIBRILLATION DETECTION BY A SEQUENTIAL HYPOTHESIS-TESTING ALGORITHM

被引:186
作者
THAKOR, NV
ZHU, YS
PAN, KY
机构
[1] Department of Biomedical Engineering, The Johns Hopkins School of Medicine, Baltimore
关键词
D O I
10.1109/10.58594
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An algorithm for detecting ventricular fibrillation (VF) and ventricular tachycardia (VT) by the method of sequential hypothesis testing is presented. The algorithm first generates a binary sequence by comparing the signal to a threshold. The probability distribution of the time intervals of the binary sequence is obtained, and Wald's sequential hypothesis testing procedure is next employed to discriminate the arrhythmias. Sequential hypothesis testing of 85 cases resulted in identification of 1) 97.64% VF and 97.65% VT episodes after 5 s, and 2) 100% identification of both VF and VT after 7 s. The desired false positive and false negative error probabilities can be preprogrammed into the algorithm. An important feature of the sequential method is that extra time for detection can be traded off for improved accuracy, and vice versa. © 1990 IEEE
引用
收藏
页码:837 / 843
页数:7
相关论文
共 18 条
[1]  
AUBERT AE, 1982, COMPUT CARDIOL, P477
[2]  
BREKELMANS FEM, 1980, P COMPUT CARDIOLL, P361
[3]   OPTIMUM SEQUENTIAL DETECTION OF SIGNALS IN NOISE [J].
BUSSGANG, JJ ;
MIDDLETON, D .
IRE TRANSACTIONS ON INFORMATION THEORY, 1955, 1 (03) :5-18
[4]   VENTRICULAR-FIBRILLATION DETECTION BY A REGRESSION TEST ON THE AUTOCORRELATION FUNCTION [J].
CHEN, S ;
THAKOR, NV ;
MOWER, MM .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1987, 25 (03) :241-249
[5]  
HANCOCK JC, 1966, SIGNAL DETECTION THE, P88
[6]  
HERBSCHLEB JN, 1979, P OCMPUT CARDIOL LON, P49
[7]  
KALBFLEISH JG, 1985, PROBABILITY STATISTI, P155
[8]  
Kuo S., 1978, Computers in Cardiology 1978, P347
[9]  
Langer A, 1976, Med Instrum, V10, P163
[10]  
MIROWSKI M, 1978, AM HEART J, V100, P996