IN-VIVO TENSILE TESTING OF FLUORAPATITE AND HYDROXYLAPATITE PLASMA-SPRAYED COATINGS

被引:97
作者
KANGASNIEMI, IMO
VERHEYEN, CCPM
VANDERVELDE, EA
DEGROOT, K
机构
[1] LEIDEN UNIV,BIOMAT RES GRP,2333 AA LEIDEN,NETHERLANDS
[2] LEIDEN UNIV,DEPT BIOSTAT,2333 AA LEIDEN,NETHERLANDS
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1994年 / 28卷 / 05期
关键词
D O I
10.1002/jbm.820280506
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A tensile test has been developed to test bioactive coating materials. Hydroxylapatite (HA) and fluorapatite (FA) coatings with various roughnesses were tested using sandblasted titanium (Ti) as a control material. Twelve goats received 7 implants each for 6 (2 goats), 12 (5), and 24 (4) week implantation periods. After 12 weeks the mean tensile strength values were highest for polished hydroxylapatite (HAP) followed by HA > FAP (polished fluorapatite) > FA > Ti, and hydroxylapatite type coatings were found to exhibit significantly higher values than fluorapatite type coatings or titanium implants. After 24 weeks no statistically significant differences could be found between any of the implant types. The order of the mean strength values was now HAP > HA > FA > FAP > Ti. Fracture always occurred between the coating and titanium if bone contact had been established. In conclusion, it is suggested that results from different types of test methods cannot be used to compare different types of bioactive coatings. (C) 1994 John Wiley and Sons, Inc.
引用
收藏
页码:563 / 572
页数:10
相关论文
共 15 条
[1]   PUSH-OUT TESTS [J].
BLACK, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1989, 23 (11) :1243-1245
[2]  
Carlsson L, 1988, Int J Oral Maxillofac Implants, V3, P21
[3]  
DEJONGE H, 1963, INLEIDING TOT MEDISC, P708
[4]   A FINITE-ELEMENT ANALYSIS OF THE PUSH-OUT TEST - INFLUENCE OF TEST CONDITIONS [J].
DHERT, WJA ;
VERHEYEN, CCPM ;
BRAAK, LH ;
DEWIJN, JR ;
KLEIN, CPAT ;
DEGROOT, K ;
ROZING, PM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (01) :119-130
[5]   A MECHANICAL INVESTIGATION OF FLUORAPATITE, MAGNESIUMWHITLOCKITE, AND HYDROXYLAPATITE PLASMA-SPRAYED COATINGS IN GOATS [J].
DHERT, WJA ;
KLEIN, CPAT ;
WOLKE, JGC ;
VANDERVELDE, EA ;
DEGROOT, K ;
ROZING, PM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (10) :1183-1200
[6]   DIFFERENCE OF BOND BONDING BEHAVIOR AMONG SURFACE-ACTIVE GLASSES AND SINTERED APATITE [J].
FUJIU, T ;
OGINO, M .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (07) :845-859
[7]  
GROSS U, 1988, QUANTITATIVE CHARACT, P211
[8]  
MATTHECK C, 1989, KFK4648 KERN FORSCH
[9]   REMOVAL TORQUE FOR BONE-CEMENT AND TITANIUM SCREWS IMPLANTED IN RABBITS [J].
MORBERG, P ;
ALBREKTSSON, T .
ACTA ORTHOPAEDICA SCANDINAVICA, 1991, 62 (06) :554-556
[10]  
OHASHI H, 1991, 16TH T ANN M SOC BIO, P264