THE PERFORMANCE OF DIFFERENT PRESSURE PULSE GENERATORS FOR EXTRACORPOREAL LITHOTRIPSY - A COMPARISON BASED ON COMMERCIAL LITHOTRIPTERS FOR KIDNEY-STONES

被引:24
作者
BUIZZA, A
DELLAQUILA, T
GIRIBONA, P
SPAGNO, C
机构
[1] CRSTBS,TRIESTE,ITALY
[2] CIVAB,ITAL TBS,TRIESTE,ITALY
[3] USL 1,DEPT BIOENGN,TRIESTE,ITALY
关键词
ESWL; SHOCK-WAVE GENERATORS; TECHNICAL EVALUATION; KIDNEY STONES; LITHOTRIPSY; ULTRASONICS;
D O I
10.1016/S0301-5629(94)00104-9
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
By using needle hydrophones and a PC-controlled experimental set-up, the acoustic output of 10 commercial extracorporeal shock-wave lithotripters has been measured. The pressure field was measured in the focus, along the beam axis, in the focal plane and ''at the skin level'' (a plane orthogonal to the beam axis, 5 cm backward from the focus, assumed as the entrance site of the pressure pulse into the patient's body). The set of tested instruments included the three technologies nowadays in use to generate the pressure pulse, namely electrohydraulic, electromagnetic and piezoelectric. Notwithstanding large intra-technology variability, the results indicate that electrohydraulic and electromagnetic generators can provide comparable performances. Piezoelectric generators produced pressure pulses with less energy than the others. Possible implications of the experimental results on treatment safety and effectiveness of kidney stone destruction are discussed.
引用
收藏
页码:259 / 272
页数:14
相关论文
共 24 条
[1]  
[Anonymous], 1992, ULTRASOUND MED BIOL, V18, P731
[2]  
BOSTROM U, 1988, SCAND J UROL NEPHR S, V122, P73
[3]   TEST FOR KIDNEY HEMORRHAGE FOLLOWING EXPOSURE TO INTENSE, PULSED ULTRASOUND [J].
CARSTENSEN, EL ;
HARTMAN, C ;
CHILD, SZ ;
COX, CA ;
MAYER, R ;
SCHENK, E .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (07) :681-685
[4]  
CATHIGNOL D, 1992, 14TH ANN INT C IEEE, P360
[5]   ACOUSTIC CAVITATION GENERATED BY AN EXTRACORPOREAL SHOCKWAVE LITHOTRIPTER [J].
COLEMAN, AJ ;
SAUNDERS, JE ;
CRUM, LA ;
DYSON, M .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (02) :69-76
[6]   ACOUSTIC-EMISSION AND SONOLUMINESCENCE DUE TO CAVITATION AT THE BEAM FOCUS OF AN ELECTROHYDRAULIC SHOCK-WAVE LITHOTRIPTER [J].
COLEMAN, AJ ;
CHOI, MJ ;
SAUNDERS, JE ;
LEIGHTON, TG .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1992, 18 (03) :267-281
[7]   PRESSURE WAVE-FORMS GENERATED BY A DORNIER EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER [J].
COLEMAN, AJ ;
SAUNDERS, JE ;
PRESTON, RC ;
BACON, DR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (10) :651-657
[8]   A SURVEY OF THE ACOUSTIC OUTPUT OF COMMERCIAL EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTERS [J].
COLEMAN, AJ ;
SAUNDERS, JE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (03) :213-227
[9]   BIOLOGICAL EFFECTS OF SHOCK-WAVES - CAVITATION BY SHOCK-WAVES IN PIGLET LIVER [J].
DELIUS, M ;
DENK, R ;
BERDING, C ;
LIEBICH, HG ;
JORDAN, M ;
BRENDEL, W .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (05) :467-472
[10]   ESTIMATION OF TRANSIENT TEMPERATURE ELEVATION IN LITHOTRIPSY AND IN ULTRASONOGRAPHY [J].
FILIPCZYNSKI, L ;
WOJCIK, J .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (07) :715-721