SERUM-CALCIUM ION ACTIVITY - SOME ASPECTS ON METHODOLOGICAL DIFFERENCES AND INTRAINDIVIDUAL VARIATION

被引:27
作者
LARSSON, L
OHMAN, S
机构
[1] Department of Clinical Chemistry, Linkoping University
关键词
D O I
10.1016/S0009-9120(79)80141-1
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Sera were analyzed for calcium ion activity on the Orion 99-20 and SS-20 systems. Samples from 84 healthy subjects were analyzed on both systems and, because of a change of electrode composition, 74 of them were reanalyzed after this change. Additional 89 were analyzed on only SS-20. The distribution of serum calcium ion activity was close to Gaussian. The difference between the 99-20 and SS-20 readings was 0.048 ± 0.017 mmol/liter (range -0.01 to 0.08) before the electrode change and -0.085 ± 0.027 mmol/liter (range -0.14 to 0.02) after. The reference ranges were 1.02-1.23 mmol/liter for 99-20 and 0.93-1.20 mmol/liter for SS-20 before the electrode change. The following characteristics may contribute to differences in the readings: higher temperature, higher flow rate, new calcium and reference electrodes, new liquid junction construction and a new pumping procedure. For 74 subjects the determination of serum calcium ion activity (99-20 system) was performed with a five year interval and the difference between the first and second measurements was -0.0046 ± 0.0168 mmol/liter (mean ± SD), ranging from -0.04 to 0.03 mmol/liter. This corresponded to an intraindividual variation of 1.5%. The corresponding variations for serum total calcium and corrected calcium were 4.8% and 3.3% respectively. © 1979.
引用
收藏
页码:138 / 141
页数:4
相关论文
共 16 条
[1]  
BUCK RP, 1978, ANAL CHEM, V50, pR17, DOI 10.1021/ac50028a003
[3]   IMPROVED PROCEDURE FOR ELECTROCHEMICAL DETERMINATION OF SERUM IONIZED CALCIUM [J].
HANSEN, AC ;
ENGEL, K ;
KILDEBERG, P ;
WAMBERG, S .
CLINICA CHIMICA ACTA, 1977, 79 (02) :507-510
[4]  
HUSDAN H, 1977, CLIN CHEM, V23, P1775
[5]  
LADENSON JH, 1973, CLIN CHEM, V19, P575
[6]  
LARSSON L, 1978, CLIN CHEM, V24, P1962
[7]  
LARSSON L, 1978, CLIN CHEM, V24, P731
[8]   ESTIMATION OF PLASMA ALBUMIN BY BCG DYE BINDING ON TECHNICON SMA 12-60ANALYSER AND A COMPARISON WITH HABA DYE BINDING TECHNIQUE [J].
LEONARD, PJ ;
PERSAUD, J ;
MOTWANI, R .
CLINICA CHIMICA ACTA, 1971, 35 (02) :409-&
[9]   POTENTIOMETRIC DETERMINATION OF SERUM IONIZED CALCIUM IN A NORMAL HUMAN POPULATION [J].
LINDGARDE, F .
CLINICA CHIMICA ACTA, 1972, 40 (02) :477-+
[10]  
LOKEN HF, 1960, J BIOL CHEM, V235, P3654