THE PB3300 INTRAARTERIAL BLOOD-GAS MONITORING-SYSTEM

被引:13
作者
LUMSDEN, T
MARSHALL, WR
DIVERS, GA
RICCITELLI, SD
机构
[1] FOxS Division, Puritan-Bennett Corporation, Carlsbad, 92009, CA
来源
JOURNAL OF CLINICAL MONITORING | 1994年 / 10卷 / 01期
关键词
MONITORING; OXYGEN; CARBON DIOXIDE; ACID-BASE EQUILIBRIUM; EQUIPMENT; SENSORS;
D O I
10.1007/BF01651467
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Continuous intraarterial blood gas (IABG) monitoring is in clinical use both in the operating room and intensive care unit. This technology uses miniature, optically-based sensors that can be placed into a patient's artery. The arterial blood gas values are transduced into an optical signal that is measured by a bedside monitor on which the values are displayed. In this paper, we describe the operating principles of the PB3300 Intra-Arterial Blood Gas Monitoring System (Puritan-Bennett Corporation, FOxS Division, Carlsbad, CA). Topics include the principles of fluorescent determinations of pH, PCO2, and PO2; the optical path of the PB3300; system calibration; dye layer geometry; and clinical operation. The accuracy, precision, and drift of the system measuring tonometered aqueous standards are reported. The following values were noted for eight sensors sending data to eight monitors: system bias and precisions of 0.00 +/- 0.02 pH at a pH of 7.40, -2.5 +/- 1.5 mm Hg PCO2 at a PCO2 of 40 mm Hg, and 3.3 +/- 1.3 mm Hg PO2 at a PO2 of 80 mm Hg.
引用
收藏
页码:59 / 66
页数:8
相关论文
共 25 条
[1]  
BARKER SJ, 1991, ANESTH ANALG, V73, P43
[2]  
BUCKLES RG, 1982, Patent No. 4321057
[3]  
CLARK LC, 1981, CRIT CARE MED, V9, P690
[4]  
CONWAY M, 1976, PEDIATRICS, V57, P224
[5]   EVALUATION OF A DUAL-FUNCTION PH AND PCO-2 INVIVO SENSOR [J].
COON, RL ;
LAI, NCJ ;
KAMPINE, JP .
JOURNAL OF APPLIED PHYSIOLOGY, 1976, 40 (04) :625-629
[6]  
FOGT EJ, 1990, CLIN CHEM, V36, P1573
[7]   OPTICAL FLUORESCENCE AND ITS APPLICATION TO AN INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM [J].
GEHRICH, JL ;
LUBBERS, DW ;
OPITZ, N ;
HANSMANN, DR ;
MILLER, WW ;
TUSA, JK ;
YAFUSO, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :117-132
[8]  
GOTTLIEB A, 1991, BIOSENSORS FIBEROPTI, P325
[9]  
Greenblott G B, 1991, J Cardiothorac Vasc Anesth, V5, P365, DOI 10.1016/1053-0770(91)90161-L
[10]  
HALBERT SA, 1990, CLIN CHEM, V36, P1581