SENSITIVE FLAME IONIZATION DETECTOR FOR THE DETERMINATION OF TRACES OF ATMOSPHERIC HYDROCARBONS BY CAPILLARY COLUMN GAS-CHROMATOGRAPHY

被引:36
作者
TSUJINO, Y
KUWATA, K
机构
[1] Environmental Pollution Control Centre, Higashinari-ku, Osaka, 537, 3-62, 1 Chome, Nakamichi
来源
JOURNAL OF CHROMATOGRAPHY | 1993年 / 642卷 / 1-2期
关键词
D O I
10.1016/0021-9673(93)80102-E
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive flame ionization detector was developed for capillary column gas chromatography. The detector used no make-up gas and lower flow-rates of hydrogen and air to suppress detector noise and to achieve maximum response. The detector was responsive down to the 10(-13) g level for C2-C-11 hydrocarbons. Atmospheric hydrocarbons could be determined at parts per billion or parts per trillion levels by using gas chromatography with the detector via cryogenic concentration of a small volume of air sample.
引用
收藏
页码:383 / 388
页数:6
相关论文
共 18 条
[1]   PRECOLUMN SAMPLE ENRICHMENT DEVICE FOR ANALYSIS OF AMBIENT VOLATILE ORGANICS BY GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
ARNTS, RR .
JOURNAL OF CHROMATOGRAPHY, 1985, 329 (03) :399-405
[2]   GAS-CHROMATOGRAPHIC EVALUATION OF THE ORGANIC-COMPONENTS PRESENT IN THE ATMOSPHERE AT TRACE LEVELS WITH THE AID OF CARBOPACK-B FOR PRECONCENTRATION OF THE SAMPLE [J].
CICCIOLI, P ;
BRANCALEONI, E ;
CECINATO, A ;
DIPALO, C ;
BRACHETTI, A ;
LIBERTI, A .
JOURNAL OF CHROMATOGRAPHY, 1986, 351 (03) :433-449
[3]  
CLERK AI, 1984, INT J ENVIRON AN CH, V17, P315
[4]   STYRENE COPOLYMERS AS POLLUTANT ADSORBENTS - SAFE SAMPLING VOLUME DETERMINATION [J].
COPPI, S ;
BETTI, A ;
ASCANELLI, M .
JOURNAL OF CHROMATOGRAPHY, 1987, 390 (02) :349-355
[5]   HYDROCARBON GASES EMITTED FROM VEHICLES ON THE ROAD .1. A QUALITATIVE GAS-CHROMATOGRAPHY MASS-SPECTROMETRY SURVEY [J].
HAMPTON, CV ;
PIERSON, WR ;
HARVEY, TM ;
UPDEGROVE, WS ;
MARANO, RS .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1982, 16 (05) :287-298
[6]   CHROMATOGRAPHIC ANALYSIS OF ORGANIC-COMPOUNDS IN THE ATMOSPHERE [J].
HUTTE, RS ;
WILLIAMS, EJ ;
STAEHELIN, J ;
HAWTHORNE, SB ;
BARKLEY, RM ;
SIEVERS, RE .
JOURNAL OF CHROMATOGRAPHY, 1984, 302 (OCT) :173-179
[7]  
HYVER KJ, 1989, HIGH RESOLUTION GAS, P6
[8]   AUTOMATED CRYOGENIC PRECONCENTRATION AND GAS-CHROMATOGRAPHIC DETERMINATION OF VOLATILE ORGANIC-COMPOUNDS IN AIR [J].
MCCLENNY, WA ;
PLEIL, JD ;
HOLDREN, MW ;
SMITH, RN .
ANALYTICAL CHEMISTRY, 1984, 56 (14) :2947-2951
[9]   NON-METHANE HYDROCARBONS IN THE ATMOSPHERE OF SYDNEY, AUSTRALIA [J].
NELSON, PF ;
QUIGLEY, SM .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1982, 16 (10) :650-655
[10]   THE HYDROCARBON COMPOSITION OF EXHAUST EMITTED FROM GASOLINE FUELED VEHICLES [J].
NELSON, PF ;
QUIGLEY, SM .
ATMOSPHERIC ENVIRONMENT, 1984, 18 (01) :79-87