DOSE DISTRIBUTION CONSIDERATIONS OF MEDIUM-ENERGY ELECTRON-BEAMS AT EXTENDED SOURCE-TO-SURFACE DISTANCE

被引:13
作者
SHAW, CB
AYYANGAR, KM
PAWLICKI, T
KORB, LJ
机构
[1] W VIRGINIA UNIV,ROBERT C BYRD HLTH SCI CTR,DIV RADIAT ONCOL,MORGANTOWN,WV 26506
[2] MED COLL OHIO,DEPT RADIAT THERAPY,TOLEDO,OH 43699
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1995年 / 32卷 / 01期
关键词
RADIATION THERAPY; EXTERNAL BEAM THERAPY; DOSIMETRY; ELECTRON BEAM; TREATMENT PLANNING;
D O I
10.1016/0360-3016(94)00598-F
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To determine the effects of extended source-to-surface distance (SSD) on dose distributions for a range of medium energy electron beams and cone sizes. Methods and Materials: The depth-dose curves and isodose distributions of 6 MeV, 10 MeV, and 14 MeV electron beams from a dual photon and multielectron energies linear accelerator were studied, To examine the influence of cone size, the smallest and the largest cone sizes available were used, Measurements were carried out in a water phantom with the water surface set at three different SSDs from 101 to 116 cm. Results: In the region between the phantom surface and the depth of maximum dose, the depth-dose decreases as the SSD increases for all electron beam energies. The effects of extended SSD in the region beyond the depth of maximum dose are unobservable and, hence, considered minimal. Extended SSD effects are apparent for higher electron beam energy with small cone size causing the depth of maximum dose and the rapid dose fall-off region to shift deeper into the phantom, However, the change in the depth-dose curve is small, On the other hand, the rapid dose fall-off region is essentially unaltered when the large cone is used, The penumbra enlarges and electron beam flatness deteriorates with increasing SSD.
引用
收藏
页码:159 / 164
页数:6
相关论文
共 10 条
[1]  
BRADY LW, 1992, PRINCIPLES PRACTICE, P246
[2]  
HO JH, 1976, HIGH ENERG PHOTONS E, P51
[3]  
HOGSTROM KR, 1984, 1ST P ANN MEV US C I, P4
[4]  
HOGSTROM KR, 1992, AAPM MED PHYSICS MON, V19, P390
[5]  
KHAN FM, 1992, LEVITT TAPLEYS TECHN, P80
[6]  
Saw C B, 1994, Med Dosim, V19, P77
[7]   COMPREHENSIVE ANALYSIS OF ELECTRON-BEAM CENTRAL AXIS DOSE FOR A RADIOTHERAPY LINEAR-ACCELERATOR [J].
SHIU, AS ;
TUNG, SS ;
NYERICK, CE ;
OCHRAN, TG ;
OTTE, VA ;
BOYER, AL ;
HOGSTROM, KR .
MEDICAL PHYSICS, 1994, 21 (04) :559-566
[8]  
TAPLE ND, 1976, CLIN APPLICATIONS EL
[9]  
1991, MED PHYS, V18, P73
[10]  
1984, ICRU35 INT COMM RAD