MEASUREMENT IN-VITRO OF HUMAN PLASMA GLYCEROL WITH A HYDROGEN-PEROXIDE DETECTING MICRODIALYSIS ENZYME ELECTRODE

被引:24
作者
MURPHY, LJ
GALLEY, PT
机构
[1] Unit of Metabolic Medicine, The Mint Wing, St. Mary's Hospital, Paddington, London
关键词
D O I
10.1021/ac00095a035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Human plasma glycerol was determined with a microdialysis electrode, containing the enzymes glycerol kinase and glycerol phosphate oxidase held stationary within the electrode. A microdialysis electrode is essentially a conventional microdialysis probe, with a platinum working electrode inserted into the tip of the dialysis fiber and reference and counter electrodes contained in the upper compartment. The linear range of response to glycerol was directly dependent on the concentration of ATP. At 4 mM ATP, the linear range was 0.5-500 mu M. A fast response time of 20 s was obtained. Two types of interferences were observed when plasma glycerol was measured: direct oxidation of interferents at the electrode and attenuation of response to glycerol by reaction with hydrogen peroxide and/or poisoning of the platinum electrode. Ascorbate, mate, and acetaminophen were removed from plasma samples by a pretreatment step involving peroxidase and catalase. Any remaining interferent current was reduced by electropolymerizing o-phenylenediamine onto the platinum electrode. Adsorption of plasma proteins on the dialysis fiber was minimal and was not reduced by the preadsorption of human serum albumin. Very good correlation was obtained between the electrode and the standard spectrophotometric technique for the variation in glycerol concentration with time.
引用
收藏
页码:4345 / 4353
页数:9
相关论文
共 29 条
  • [1] AEBI HE, 1974, METHOD ENZYMAT AN, V2, P267
  • [2] THE DIALYSIS ELECTRODE - A NEW METHOD FOR INVIVO MONITORING
    ALBERY, WJ
    BOUTELLE, MG
    GALLEY, PT
    [J]. JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1992, (12) : 900 - 901
  • [3] A DIALYSIS ELECTRODE FOR GLYCEROL
    ALBERY, WJ
    GALLEY, PT
    MURPHY, LJ
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1993, 344 (1-2): : 161 - 166
  • [4] AMPEROMETRIC ENZYME ELECTRODES .1. THEORY
    ALBERY, WJ
    BARTLETT, PN
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1985, 194 (02): : 211 - 222
  • [5] ALTMAN P, 1971, BLOOD OTHER BODY FLU
  • [6] BRETT AMO, 1993, BIOELECTROCH BIOENER, V31, P311
  • [7] POTENTIOMETRIC QUANTITATION OF GLYCEROL USING IMMOBILIZED GLYCEROL DEHYDROGENASE
    CHEN, AK
    STARZMANN, JA
    LIU, CC
    [J]. BIOTECHNOLOGY AND BIOENGINEERING, 1982, 24 (04) : 971 - 975
  • [8] DAVIS G, 1984, Patent No. 125867
  • [9] DEFRONZO RA, 1979, AM J PHYSIOL, V237, pE214
  • [10] ESDERS TW, 1979, J BIOL CHEM, V254, P2710