PERFORMANCE-CHARACTERISTICS AND CLINICAL-EVALUATION OF AN IN-VITRO BLEEDING-TIME DEVICE - THROMBOSTAT-4000

被引:8
作者
ALSHAMEERI, RS
MAMMEN, EF
机构
[1] WAYNE STATE UNIV,SCH MED,DEPT PATHOL,DETROIT,MI
[2] WAYNE STATE UNIV,SCH MED,DEPT PHYSIOL,DETROIT,MI
[3] WAYNE STATE UNIV,SCH MED,DEPT OBSTET & GYNECOL,DETROIT,MI
关键词
PLATELET FUNCTION TESTS; BLEEDING TIME; BLOOD PLATELET DISORDERS; VON WILLEBRANDS DISEASE; THROMBOSTAT;
D O I
10.1016/0049-3848(95)00114-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The performance characteristics of an in vitro bleeding time device - Thrombostat 4000 were evaluated and compared with the Simplate bleeding time in healthy individuals and patients with disorders of primary hemostasis. Reference ranges were established using 30 normal volunteers. Although there were variations between different filter batches, reproducibility was good within a single batch. There were no differences between the two channels of the instrument and between male and female subjects. Hematocrit correlated negatively with the initial flow (IF) and IF correlated positively with closure time (T) and bleeding volume (V). Aspirin could be detected only when the traditional addition of ADP was replaced with CaCl2. Both, closure time (T) or bleeding volume (V) were more sensitive than Simplate bleeding time and T was more sensitive than V in detecting patients with disorders of primary hemostasis. We conclude that the Thrombostat 4000 is a reproducible, reliable, sensitive and easy to use instrument, It is superior to the traditional in vivo bleeding times for investigations of disorders of primary hemostasis (screening, diagnosis, monitoring, etc.).
引用
收藏
页码:275 / 287
页数:13
相关论文
共 24 条
[1]  
ALSHAMEERI R, 1993, THROMB HAEMOSTASIS, V69, P1146
[2]  
BICK RL, 1993, HEMATOLOGY CLIN LABO, V2, P1369
[3]   PLATELET-FUNCTION STUDIES IN BERNARD-SOULIER SYNDROME [J].
BITHELL, TC ;
STRONG, RR ;
PAREKH, SJ .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1972, 201 (OCT27) :145-&
[4]  
DIETRICH G, 1990, INFUSIONSTHERAPIE, V17, P214
[5]  
DIETRICH G, 1990, INFUSIONSTHERAPIE, V17, P212
[6]   The relation of blood platelets to hemorrhagic disease - Description of a method for determining the bleeding time and coagulation time and report of three cases of hemorrhagic disease relieved by transfusion [J].
Duke, WW .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1910, 55 :1185-1192
[7]   HEMOSTATOMETER - A NEW INVITRO TECHNIQUE FOR ASSESSING HEMOSTATIC ACTIVITY OF BLOOD [J].
GOROG, P ;
AHMED, A .
THROMBOSIS RESEARCH, 1984, 34 (04) :341-357
[8]   PLATELET STORAGE POOL DEFICIENCY - DIAGNOSIS IN PATIENTS WITH PROLONGED BLEEDING TIMES AND NORMAL PLATELET-AGGREGATION [J].
ISRAELS, SJ ;
MCNICOL, A ;
ROBERTSON, C ;
GERRARD, JM .
BRITISH JOURNAL OF HAEMATOLOGY, 1990, 75 (01) :118-121
[9]  
Ivy AC, 1941, J LAB CLIN MED, V26, P1812
[10]  
KRATZER MAA, 1985, HAEMOSTASIS, V15, P357