DETECTOR, SHIELDING AND GEOMETRIC DESIGN FACTORS FOR A HIGH-RESOLUTION PET SYSTEM

被引:27
作者
DIGBY, WM
DAHLBOM, M
HOFFMAN, EJ
机构
[1] UNIV CALIF LOS ANGELES,NUCL MED LAB,LOS ANGELES,CA 90024
[2] UNIV CALIF LOS ANGELES,BIOMED & ENVIRONM SCI LAB,LOS ANGELES,CA 90024
关键词
D O I
10.1109/23.106695
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We evaluated the resolution, efficiency and scatter rejection on a new high resolution PET system designed for animal studies which is based on a 2-D modular detector system. A digital positioning system was evaluated by testing different encoding methods. Tungsten inter-plane septa of different thicknesses and geometries were evaluated by Monte Carlo simulations and experiments. The detector system consists of a 6×8 array of BGO crystals coupled to 2 dual photomultiplier tubes (PMTs). The crystals are 3.5 mm wide with 4 mm spacing transaxially and are 6.25 mm long with 6.75 mm spacing axially. PMT outputs are digitized and Anger camera type logic is used to determine the X and Y location of the scintillation event. Data is then sorted into the appropriate detector using a column look up table then column dependent row look up tables. The digital nature of the data collection system, which allows optimization of crystal identification, was found to be essential to the performance of the detector system. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:664 / 670
页数:7
相关论文
共 8 条
[1]  
BOHM C, 1986, SPIE P, V671, P254
[2]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[3]   AN EVALUATION OF A TWO-DIMENSIONAL ARRAY DETECTOR FOR HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (04) :264-272
[4]   A STUDY OF THE POSSIBILITY OF USING MULTI-SLICE PET SYSTEMS FOR 3D IMAGING [J].
DAHLBOM, M ;
ERIKSSON, L ;
ROSENQVIST, G ;
BOHM, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :1066-1071
[5]  
HOFFMAN E J, 1988, Journal of Nuclear Medicine, V29, P983
[6]  
HOFFMAN EJ, 1987, J NUCL MED, V28, P758
[7]   A QUAD BGO DETECTOR AND ITS TIMING AND POSITIONING DISCRIMINATION FOR POSITRON COMPUTED-TOMOGRAPHY [J].
MURAYAMA, H ;
NOHARA, N ;
TANAKA, E ;
HAYASHI, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 192 (2-3) :501-511
[8]   EMISSION COMPUTED-TOMOGRAPHY SIMULATOR [J].
RICCI, AR ;
HOFFMAN, EJ ;
PHELPS, ME ;
LOW, GC ;
KUHL, DE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1980, 27 (01) :479-484