摘要
When a low-nanogram amount of N1,N3-bis-(pentafluorobenzyl)-N7-(2-[pentafluorobenzyloxy]ethyl)xanthine was subjected to HPLC, low-picogram amounts of the compound could be detected subsequently (off-line by gas chromatography-electron-capture negative-ion mass spectrometry) after injection of pure mobile phase. This was in spite of significant, intermediate washing of the injector and column. It was determined that essentially 99.9% of this analyte contamination came from the injector. Use of two injectors is a practical remedy for this problem.
引用
收藏
页码:161 / 163
页数:3
相关论文
共 5 条
[1]  
BAKALYAR S, COMMUNICATION
[2]  
DOLAN JW, 1991, LC GC-MAG SEP SCI, V9, P22
[3]  
LIN D, 1987, J CHROMATOGR, V422, P61
[4]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY VERSUS SOLID-PHASE EXTRACTION FOR POST-DERIVATIZATION CLEANUP PRIOR TO GAS-CHROMATOGRAPHY ELECTRON-CAPTURE NEGATIVE-ION MASS-SPECTROMETRY OF N1,N3-BIS-(PENTAFLUOROBENZYL)-N7-(2-[PENTAFLUOROBENZYLOXY]ETHYL)XANTHINE, A PRODUCT DERIVED FROM AN ETHYLENE-OXIDE DNA ADDUCT [J].