ENDOSCOPIC PROTECTOR HOOD FOR SAFE REMOVAL OF SHARP-POINTED GASTROESOPHAGEAL FOREIGN-BODIES

被引:18
作者
BERTONI, G
PACCHIONE, D
CONIGLIARO, R
SASSATELLI, R
PEDRAZZOLI, C
BEDOGNI, G
机构
[1] Department of Digestive Endoscopy, S. Maria Nuova Hospital, Reggio Emilia, I-42100
来源
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES | 1992年 / 6卷 / 05期
关键词
SHARP-POINTED FOREIGN BODIES; UPPER GASTROINTESTINAL TRACT; ENDOSCOPIC REMOVAL; PROTECTOR HOOD;
D O I
10.1007/BF02498816
中图分类号
R61 [外科手术学];
学科分类号
摘要
Endoscopic removal of sharp and pointed foreign bodies in the upper gastrointestinal tract still poses technical difficulties. Overtubes may be uncomfortable to the patients and only objects fitting within a lumen of about 11-15 mm can be removed. We present here our preliminary experience with a commercially available new endoscopic end protector hood having the unique characteristic of maintaining its bell portion inverted during the progression of the scope through the gut and flipping back to its original shape during withdrawal through the lower esophageal sphincter. This prevents exposure of the esophageal and pharyngeal wall to injuries from the edges of the foreign body. We used this device to remove nine large sharp or pointed gastroesophageal foreign bodies in six consecutive patients in whom conventional techniques were considered unsafe or not technically feasible owing to the size and shape of the objects. In all cases, removal was successful and easily performed without bowel damage. We believe that this device enhances the safe removal of a wide variety of potentially dangerous gastroesophageal foreign bodies.
引用
收藏
页码:255 / 258
页数:4
相关论文
共 12 条
[1]  
Bedogni G., Oselladore D., Ricci E., Rossoni R., Operative endoscopy of the digestive tract, (1984)
[2]  
Carp L., Foreign bodies in the intestine, Annals of Surgery, 85, pp. 575-559, (1927)
[3]  
Garrido J., Barkin J.S., Endoscopic modification for safe foreign body removal, Am J Gastroenterol, 80, pp. 957-958, (1985)
[4]  
Martin W.C., Jones S., Noriyaschi N., Texter E.C., Fiberendoscopic removal of long foreign bodies from the stomach using a colon stiffening tube as a protective sheath, Am J Gastroenterol, 71, pp. 505-507, (1979)
[5]  
Mee A.S., Wright J.P., Endoscopic removal of sharp foreign objects using an oversleeve, S Afr Med J, 60, pp. 743-744, (1981)
[6]  
Clemente Ricote G., Robago Torre L., Perez de Ayala V., Castellanos D., Menchen P., Senent C., Velo J.L., Robles J., Alcata-Santaella R., Fiberendoscopic removal of foreign bodies of the upper gastrointestinal tract, Surg Gyn Obst, 160, pp. 499-504, (1985)
[7]  
Rogers B.H.G., Kot C., Meiri S., Epstein M., An overtube for the flexible fiberoptic esophagogastroduodenoscope, Gastrointest Endosc, 28, pp. 256-257, (1982)
[8]  
Rosch W., Classen M., Fiberendoscopic foreign body removal from the upper gastrointestinal tract, Endoscopy, 4, pp. 193-197, (1972)
[9]  
Spurling T.J., Zaloga G.P., Richter J.E., Fiberendoscopic removal of a gastric foreign body with overtube technique, Gastrointest Endosc, 29, pp. 226-227, (1983)
[10]  
Sykes M.T., Schwesinger W.H., Safe retrieval of a swallowed denture using a hooded panendoscope, Gastrointest Endosc, 29, (1983)