AUTOMATED-DETERMINATION OF TOTAL PLASMA-CHOLESTEROL - SERUM CALIBRATION TECHNIQUE

被引:34
作者
BACHORIK, PS
WOOD, PDS
WILLIAMS, J
KUCHMAK, M
AHMED, S
LIPPEL, K
ALBERS, J
机构
[1] STANFORD UNIV,SCH MED,DEPT MED,STANFORD,CA 94305
[2] CTR DIS CONTROL,LIPIDS SECT,ATLANTA,GA 30333
[3] UNIV WASHINGTON,SCH MED,DEPT BIOCHEM,SEATTLE,WA 98195
[4] GEORGETOWN UNIV,SCH MED,WASHINGTON,DC 20007
[5] NHLBI,BETHESDA,MD 20014
[6] UNIV WASHINGTON,SCH MED,DEPT MED,SEATTLE,WA 98195
关键词
D O I
10.1016/0009-8981(79)90064-0
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The automated (AutoAnalyzer II) determination of cholesterol in nine serum pools in the concentration range 3.465-8.871 mmol/1, gave results that were approximately 10% higher than reference values when the analyses were based on unesterified cholesterol standards containing the same amount of water as the sample extracts (1963 analyses in 12 laboratories during a 12 month period; automated value = 0.032 + 1.10 × (Reference value)). A serum calibration procedure was successful in correcting the observed serum pool values to within +0.2 to -2.1% of reference values, and was equally effective in correcting the values observed for aliquots of 368 fresh-frozen plasma samples analyzed in each of the 12 laboratories during a 38 month period. © 1979.
引用
收藏
页码:145 / 153
页数:9
相关论文
共 13 条
  • [1] ABELL LL, 1952, J BIOL CHEM, V195, P357
  • [2] Bachorik P. S., 1978, Protides of the biological fluids, P307
  • [3] BACHORIK PS, 1977, HYPERLIPIDEMIA DIAGN, P41
  • [4] FREDRICKSON DS, 1972, METABOLIC BASIS INHE, P545
  • [5] GJONE E, 1978, METABOLIC BASIS INHE, P589
  • [6] KUCHMAK M, Patent No. 3764556
  • [7] LAROSA JC, 1977, HYPERLIPIDEMIA DIAGN, P205
  • [8] Tonks D., 1967, CLIN BIOCHEM, V1, P12, DOI [10.1016/S0009-9120(67)80002-X, DOI 10.1016/S0009-9120(67)80002-X]
  • [10] WILLIAMS J, 1973, CLIN CHEM, V19, P650