A cumulative sum technique has been specially designed to monitor the error between replicate determinations made on quality control plasma for consecutive batches of assays. This procedure has played a vital role in assessing assay performance. Special consideration has been given to small samples sizes (n=2 or 3) which is generally the rule rather than the exception in many situations. This technique has been applied to numerous steroid radioimmunoassays and has ensured that both the mean value and the standard error of hormone levels of a quality control pool were under control. Data from routine assays of oestriol and testosterone in plasma from women are presented. Since this technique provides a sensitive measure of monitoring error, it assists the endocrinologist in elucidating statistical inferences which are a manifestation of assay performance.
引用
收藏
页码:365 / 372
页数:8
相关论文
共 17 条
[1]  
BARNARD GA, 1959, J ROY STAT SOC B, V21, P239
[2]  
BUTTNER J, 1978, CLIN CHEM, V24, P1213
[3]   QUALITY-CONTROL IN A RADIOIMMUNOASSAY - OBSERVATIONS ON OPERATION OF A SEMIAUTOMATED ASSAY FOR HUMAN PLACENTAL LACTOGEN [J].
CHALLAND, GS ;
CHARD, T .
CLINICA CHIMICA ACTA, 1973, 46 (02) :133-138
[4]  
DUNCAN AJ, 1974, QUALITY CONTROL IND, P464
[5]   SAMPLING INSPECTION OF CONTINUOUS PROCESSES WITH NO AUTOCORRELATION BETWEEN SUCCESSIVE RESULTS [J].
EWAN, WD ;
KEMP, KW .
BIOMETRIKA, 1960, 47 (3-4) :363-380
[6]  
KEMP KW, 1961, J ROY STAT SOC B, V23, P149
[7]  
KEMP KW, 1967, J ROY STAT SOC B, V29, P248
[8]   COMPUTERIZED APPROACH TO STATISTICAL QUALITY-CONTROL FOR RADIOIMMUNOASSAYS IN CLINICAL-CHEMISTRY LABORATORY [J].
MCDONAGH, BF ;
MUNSON, PJ ;
RODBARD, D .
COMPUTER PROGRAMS IN BIOMEDICINE, 1977, 7 (03) :179-190
[9]  
RODBARD D, 1974, CLIN CHEM, V20, P1255
[10]  
Rodbard D, 1973, RADIOIMMUNOASSAY REL, V1, P165