THERAPY IMAGING - A SIGNAL-TO-NOISE ANALYSIS OF A FLUOROSCOPIC IMAGING-SYSTEM FOR RADIOTHERAPY LOCALIZATION

被引:49
作者
MUNRO, P
RAWLINSON, JA
FENSTER, A
机构
[1] ONTARIO CANC INST,TORONTO M4X 1K9,ONTARIO,CANADA
[2] UNIV WESTERN ONTARIO,DEPT RADIOL & BIOPHYS,LONDON N6A 3K7,ONTARIO,CANADA
[3] ROBARTS RES INST,LONDON N6A 5K8,ONTARIO,CANADA
关键词
fluorescent screens; fluoroscopic portal imaging systems; localization; on-line portal imaging; radiotherapy; radiotherapy verification; TV cameras;
D O I
10.1118/1.596474
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have been developing a digital fluoroscopic imaging system to replace the portal films that are currently used to verify patient positioning during radiotherapy treatments. Our system differs from previously reported devices in the construction of the detector and in the operation of the TV camera. The signal, noise, and signal-to-noise properties of this system have been determined by measuring the modulation transfer function [MTF (f)], the noise power spectra [NPS (f)], and by calculating the detective quantum efficiency [DQE(f)] of the system. The results show: (i) that the spatial resolution of the system is determined largely by the lens of the TV camera and by frame grabber; and (ii) that the noise in the system is dominated by the secondary light quanta, due to the poor light collection efficiency of the optical chain. Despite these physical limitations, a contrast-detail study shows that the fluoroscopic system is better at detecting large, low contrast objects than portal films. Therefore the system is already a reasonable alternative to portal films and modifications to the metal plate/phosphor detector, lens, TV camera, and frame grabber should improve the performance of the system further. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:763 / 772
页数:10
相关论文
共 22 条
  • [1] BAILY NA, 1980, INT J RADIAT ONCOL, V6, P935, DOI 10.1016/0360-3016(80)90341-7
  • [2] CHARACTERIZATION OF FLUORESCENT SCREENS FOR IMAGING APPLICATIONS WITH MEV NEUTRONS AND PHOTONS
    BERZINS, GJ
    LUMPKIN, AH
    SMITH, HL
    [J]. OPTICAL ENGINEERING, 1983, 22 (05) : 633 - 642
  • [3] METAL-PHOSPHOR INTENSIFYING SCREENS FOR HIGH-ENERGY IMAGING APPLICATIONS
    BUCHANAN, RA
    SKLENSKY, AF
    MAPLE, TG
    BAILEY, HN
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1974, NS21 (01) : 692 - 694
  • [4] EFFICIENCY OF HUMAN VISUAL SIGNAL DISCRIMINATION
    BURGESS, AE
    WAGNER, RF
    JENNINGS, RJ
    BARLOW, HB
    [J]. SCIENCE, 1981, 214 (4516) : 93 - 94
  • [5] CHENG YA, 1988, PHYS MED BIOL S1, V33, P46
  • [6] COHEN G, 1985, AAPM MED PHYSICS MON, V12, P141
  • [7] A METHOD FOR MODULATION TRANSFER-FUNCTION DETERMINATION FROM EDGE PROFILES WITH CORRECTION FOR FINITE-ELEMENT DIFFERENTIATION
    CUNNINGHAM, IA
    FENSTER, A
    [J]. MEDICAL PHYSICS, 1987, 14 (04) : 533 - 537
  • [8] DOI K, 1982, HHS FDA828187 PUBL
  • [9] INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .6. MTFS OF II-TV DIGITAL IMAGING-SYSTEMS
    FUJITA, H
    DOI, K
    GIGER, ML
    [J]. MEDICAL PHYSICS, 1985, 12 (06) : 713 - 720
  • [10] DIGITAL IMAGE-PROCESSING - EFFECT ON DETECTABILITY OF SIMULATED LOW-CONTRAST RADIOGRAPHIC PATTERNS
    ISHIDA, M
    DOI, K
    LOO, LN
    METZ, CE
    LEHR, JL
    [J]. RADIOLOGY, 1984, 150 (02) : 569 - 575