COMPLEX BIOELECTRIC IMPEDANCE MEASUREMENT SYSTEM FOR THE FREQUENCY-RANGE FROM 5 HZ TO 1 MHZ

被引:92
作者
ACKMANN, JJ [1 ]
机构
[1] DEPT VET AFFAIRS MED CTR,MILWAUKEE,WI
关键词
ELECTRIC CONDUCTIVITY; ELECTRONIC INSTRUMENTATION;
D O I
10.1007/BF02367609
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Analytic techniques that have been successfully employed in materials science, and to a lesser extent in the study of biologic systems, have potential for improving the application of bioelectric impedance provided that both real and imaginary impedance components can be measured with sufficient accuracy over a given frequency range. Since biologic tissue, particularly animal tissue, is typically highly conductive, phase angles are small, making accurate measurements difficult. A practical four-terminal system employing commercial lock-in amplifiers is described and error sources and corrective techniques are discussed.
引用
收藏
页码:135 / 146
页数:12
相关论文
共 10 条
  • [1] ACKMANN JJ, 1984, CRIT REV BIOMED ENG, V11, P281
  • [2] CORNETET WH, 1975, ELECTRONIC CIRCUITS
  • [3] Geddes L. A., 1972, ELECTRODES MEASUREME
  • [4] GEDDES LA, 1975, PRINCIPLES APPLIED B
  • [5] GRAEME JG, 1973, APPLICATIONS OPERATI
  • [6] KING MA, 1978, THESIS MARQUETTE U M
  • [7] Macdonald J.R., 1987, IMPEDANCE SPECTROSCO
  • [8] Millman J., 1972, INTEGRATED ELECTRONI
  • [9] SMITH GA, 1982, THESIS MARQUETTE U M
  • [10] REDUCTION OF INTERFERENCE DUE TO COMMON-MODE VOLTAGE IN BIOPOTENTIAL AMPLIFIERS
    WINTER, BB
    WEBSTER, JG
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (01) : 58 - 62