RAPID PARTIAL DIGESTION OF BIOLOGICAL TISSUES WITH NITRIC-ACID FOR THE DETERMINATION OF TRACE-ELEMENTS BY ATOMIC SPECTROMETRY

被引:43
作者
NIAZI, SB
LITTLEJOHN, D
HALLS, DJ
机构
[1] UNIV STRATHCLYDE,DEPT PURE & APPL CHEM,THOMAS GRAHAM BLDG,CATHEDRAL ST,GLASGOW G1 1XL,SCOTLAND
[2] GLASGOW ROYAL INFIRM,INST BIOCHEM,TRACE ELEMENT UNIT,GLASGOW G4 0SF,SCOTLAND
关键词
WET DIGESTION; BIOLOGICAL TISSUE; ATOMIC ABSORPTION SPECTROMETRY; INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROMETRY; TRACE ELEMENTS;
D O I
10.1039/an9931800821
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Determination of Cu, Fe, Mn and Zn in fresh bovine liver and BCR bovine liver and pig kidney reference materials, after digestion at 105-degrees-C with nitric acid for various times, showed that the trace elements were completely released after heating for only 20 min. Centrifugation of the sameples after digestion improved the separation from undigested fat. This partial-digestion method, based on heating about 0.2 g of sample with 2 ml of nitric acid for 20 min, was tested on a range of biological reference materials and on fresh bovine liver for the four elements. Comparison was made with results obtained after digestion by a recommended hydrogen peroxide-sulfuric acid digestion procedure. Results obtained by inductively coupled plasma atomic emission spectrometry or by flame and graphite-furnace atomic absorption spectrometry showed good agreement between the two digestion procedures, and results for the reference materials agreed well with the certified values. Between-batch precision was better than 4% for Cu, Fe and Zn at levels of 30 to 190,210 to 320 and 100 to 140 mug g-1, respectively. For Mn, the precision varied between 7 and 14% for measured concentrations of 8 to 12 mug g-1. The partial-digestion procedure offers simplicity, speed, low cost and the ability to handle a large number of samples at the same time.
引用
收藏
页码:821 / 825
页数:5
相关论文
共 17 条
[1]  
[Anonymous], 1970, DESTRUCTION ORGANIC
[2]   ELEMENTAL ANALYSIS OF BOVINE LIVER BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY BY USING A SIMPLE DISSOLUTION PROCEDURE [J].
ASP, TN ;
LUND, W .
TALANTA, 1992, 39 (05) :563-566
[3]   INVESTIGATION INTO MORE RAPID PROCEDURES FOR PREPARING LEAF-TISSUE SAMPLES FOR ANALYSIS [J].
BASSON, WD ;
BOHMER, RG .
ANALYST, 1972, 97 (1155) :482-&
[4]   SOLID SAMPLING IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING COMMERCIAL ATOMIZERS - A REVIEW [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (05) :353-374
[5]   A COMPARATIVE-EXAMINATION OF SAMPLE TREATMENT PROCEDURES FOR ICAP-AES ANALYSIS OF BIOLOGICAL TISSUE [J].
DEBOER, JLM ;
MAESSEN, FJMJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1983, 38 (5-6) :739-746
[6]  
HALLS D J, 1988, Analytical Proceedings, V25, P232
[7]   SPEEDING UP DETERMINATIONS BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
HALLS, DJ .
ANALYST, 1984, 109 (08) :1081-1084
[8]  
HALLS DJ, 1985, UNPUB
[9]   ATOMIC-ABSORPTION ANALYSIS OF LIVER WITHOUT ASHING [J].
HINNERS, TA .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1975, 277 (05) :377-378
[10]   MICROWAVE-ENERGY FOR ACID DECOMPOSITION AT ELEVATED-TEMPERATURES AND PRESSURES USING BIOLOGICAL AND BOTANICAL SAMPLES [J].
KINGSTON, HM ;
JASSIE, LB .
ANALYTICAL CHEMISTRY, 1986, 58 (12) :2534-2541