CONSTANT HEAT-FLUX CALORIMETER FOR MEASURING HEAT-TRANSFER DATA

被引:11
作者
IGARASHI, S
KATAYAMA, S
机构
[1] Materials Science Laboratory, Research Department, Toray Research Center, Inc., Otsu, 520, Sonoyama
关键词
D O I
10.1016/0014-3057(79)90035-1
中图分类号
O63 [高分子化学(高聚物)];
学科分类号
070305 ; 080501 ; 081704 ;
摘要
A new isothermal calorimeter, designed especially for simultaneously measuring the heat of reaction and the heat transfer data in a reaction solution, is described. The system, called a Constant Heat-Flux Calorimeter, is similar to a differential scanning calorimeter in terms of direct calorimetric measurement of the energies of reaction, but differs from the conventional calorimeters described in the literature. With this device, one recorded output in a temperature control circuit is a linear function of change in energies of reaction, and second in a differential temperature control circuit is found to be proportional to a resistance to heat transfer in a solution. The performance of the calorimeter was evaluated on the basis of some results on heat transfer data of aqueous solution of polyethylene glycol and on solution polymerization of styrene at constant temperatures. © 1979.
引用
收藏
页码:805 / 809
页数:5
相关论文