ALKALINE ESOPHAGEAL REFLUX - AN ARTIFACT DUE TO OXYGEN CORROSION OF ANTIMONY PH ELECTRODES

被引:18
作者
SJOBERG, F [1 ]
GUSTAFSSON, U [1 ]
TIBBLING, L [1 ]
机构
[1] LINKOPING UNIV HOSP, DEPT OTORHINOLARYNGOL, S-58185 LINKOPING, SWEDEN
关键词
ALKALINE REFLUX; ANTIMONY ELECTRODES; GLASS ELECTRODES; ESOPHAGUS; PH; 24-H PH MONITORING;
D O I
10.3109/00365529209028142
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Antimony electrodes are widely used for gastro-oesophageal pH monitoring. They are also sensitive to oxygen, however, especially at low PO2 levels, which are known to shift recorded values in the alkaline direction. This study, which compares antimony and glass electrodes for oesophageal pH monitoring in six adults, shows that values recorded by antimony electrodes are 2.1 +/- 0.8 pH units (mean +/- SD) higher than by glass electrodes (p < 0.001; n = 7642). A further 52 patients with suspected gastro-oesophageal reflux were investigated by 24-h pH monitoring by means of antimony electrodes. In these patients the oesophageal pH was higher than 8.0 for 7% of the time (range, 0-60%). The alkaline periods recorded with antimony electrodes were all protracted in time, smoothly increasing from a neutral pH, and did not correspond to a sudden increase in pH, which would be expected if alkaline reflux had occurred. It is concluded that high pH values obtained by antimony electrodes are due to the oxygen sensitivity of the electrodes. The diagnosis of alkaline reflux seems to be valid only when pH monitoring is performed with glass electrodes or when values obtained with antimony electrodes are adjusted for the influence of the oxygen tension in the oesophagus.
引用
收藏
页码:1084 / 1088
页数:5
相关论文
共 23 条
[1]   COMBINED PH AND PRESSURE MEASUREMENT DEVICE FOR ESOPHAGEAL INVESTIGATIONS [J].
ASK, P ;
EDWALL, G ;
TIBBLING, L .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1981, 19 (04) :443-446
[2]   ACCURACY AND CHOICE OF PROCEDURES IN 24-HOUR ESOPHAGEAL PH MONITORING WITH MONOCRYSTALLINE ANTIMONY ELECTRODES [J].
ASK, P ;
EDWALL, G ;
JOHANSSON, KE .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1986, 24 (06) :602-608
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
BUYTENDIJK FJJ, 1927, ARCH NEERLAND PHYSL, V12, P319
[5]   IMPROVED ANTIMONY-ANTIMONY(III) OXIDE PH ELECTRODES [J].
EDWALL, G .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1978, 16 (06) :661-669
[6]   PROSPECTIVE TRIAL COMPARING A COMBINATION PH PROBE NASOGASTRIC TUBE WITH ASPIRATED GASTRIC PH IN INTENSIVE-CARE UNIT PATIENTS [J].
EISENBERG, PG ;
CORT, D ;
ZUCKERMAN, GR .
CRITICAL CARE MEDICINE, 1990, 18 (10) :1092-1095
[7]   24-HOUR ESOPHAGEAL 2-LEVEL PH MONITORING IN HEALTHY-CHILDREN AND ADOLESCENTS [J].
GUSTAFSSON, PM ;
TIBBLING, L .
SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1988, 23 (01) :91-94
[8]  
HOLMQVIST A, 1934, SV KEM TIDSKR, V46, P2
[9]   ESOPHAGEAL REFLUX TESTS, MANOMETRY, ENDOSCOPY, BIOPSY, AND RADIOLOGY IN HEALTHY-SUBJECTS [J].
JOHANSSON, KE ;
BOERYD, B ;
FRANSSON, SG ;
TIBBLING, L .
SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1986, 21 (04) :399-406
[10]  
JOHNSON LF, 1974, AM J GASTROENTEROL, V62, P325