FACTORS INFLUENCING BETWEEN-LABORATORY VARIABILITY OF C-REACTIVE PROTEIN RESULTS AS EVIDENCED BY THE BELGIAN EXTERNAL QUALITY ASSESSMENT (EQA) SCHEME

被引:9
作者
DEVLEESCHOUWER, N
LIBEER, JC
CHAPELLE, JP
STRUWAY, CL
GYSSELS, C
LHOIR, A
ALBERT, A
机构
[1] UNIV LIEGE,SERV CHIM MED,LIEGE,BELGIUM
[2] UNIV LIEGE,SERV INFORMAT & STAT MED,LIEGE,BELGIUM
关键词
ACUTE PHASE PROTEINS; CRP; EXTERNAL QUALITY ASSESSMENT;
D O I
10.3109/00365519409085467
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Based on results from the Belgian External Quality Assessment (EQA) Scheme, we studied the main factors affecting the between-laboratory variation of C-reactive protein determination. Participants using homogeneous systems with several calibration points generally achieved better performance. Working temperatures influenced the results to a lesser extent. The present study stresses the importance for EQA organizers to collect more detailed information about CRP analytical methods used by the participants. It also suggests that manufacturers should be more involved in the management of quality, in particular by striving for standardization of the material (kit and calibrator) they produce for CRP assay.
引用
收藏
页码:435 / 440
页数:6
相关论文
共 9 条
[1]  
BACHNER P, 1992, DIS MARKERS, V10, P27
[2]  
FRASER CG, 1992, EUR J CLIN CHEM CLIN, V30, P311
[3]  
FRASER CG, 1993, CLIN CHEM, V39, P1447
[4]  
GALLOU G, 1993, CLIN CHEM, V39, P918
[5]  
LAUDER I, 1991, UK NATIONAL EXTERNAL
[6]   INTERFERENCE OF IGM RHEUMATOID-FACTOR WITH NEPHELOMETRIC C-REACTIVE PROTEIN DETERMINATIONS [J].
MULLER, W ;
MIERAU, R ;
WOHLTMANN, D .
JOURNAL OF IMMUNOLOGICAL METHODS, 1985, 80 (01) :77-90
[7]   FALSE RISE IN C-REACTIVE PROTEIN IN A PATIENT WITH MONOCLONAL IGM IMMUNOGLOBULIN [J].
PONGE, TD ;
LECARRER, DL ;
SAGNIEZ, MM ;
PONTET, M ;
COTTIN, SL .
CLINICA CHIMICA ACTA, 1993, 220 (01) :101-106
[8]  
Tukey JW, 1977, EXPLORATORY DATA ANA
[9]  
1987, WHO TECHNICAL REV SE, V760, P22